background image

Instructions 

DPG1000DR, F4DR

Types of Gauges

Gauge reference types read zero with the gauge port open. 
Bipolar ranges read positive pressure and vacuum in the same 
units, and zero with the gauge port open.
1000 psi and higher sensor are a sealed reference type. They read 
zero with the gauge port open are internally referenced to 14.7 psi. 
They are functionally similar to gauge reference sensors.
Absolute reference gauges read zero at full vacuum and atmospheric 
pressure with the gauge port open. With an open gauge port, read-
ings will vary continuously due to fluctuating barometric pressure.

Calibration Preparation

Gauges are calibrated at the factory using equipment traceable to NIST. 
There is no need to calibrate the gauge before putting it into service. 
Calibration should only be performed by qualified individuals using 
appropriate calibration standards and procedures. Gauges can 
be returned to factory for certified recalibration and repairs. NIST 
traceability is available.
Calibration intervals depend on your quality control program 
requirements and as-found data. Many customers calibrate their 
equipment annually.
The calibration equipment should be at least four times more accu-
rate than the gauge being calibrated. The calibration system must 
be able to generate and measure pressure and/or vacuum over the 
full range of the gauge. 
A vacuum pump able to produce a vacuum of 100 microns (0.1 
torr or 100 millitorr) or lower is required for vacuum and absolute 
gauges.
Use a stable DC power supply and an accurate mA meter for 
calibration of 4-20 mA transmitters and an accurate volt meter for 
calibration of voltage transmitters.
Allow the gauge to equalize to normal room temperature for at least 
20 minutes before calibration.

Calibration

NEVER connect retransmission output wires together or to an 
external power source or permanent damage not covered by war-
ranty will result.
For 4-20 mA output models, be sure to observe the output compli-
ance (voltage drive) capabilities of the gauge. The compliance, and 
therefore the maximum loop resistance the output can drive, is a 
function of the supply voltage to the gauge. 
Consult the graph shown below for maximum loop resistance vs. power 
supply voltage. Too large a loop resistance will cause the gauge output 
to “limit” or saturate before reaching its full 20 mA output.
When using the 0-2 volt retransmission output, do not allow the 
resistive load on the output to fall below 5K ohms. Avoid large 
capacitive loads (greater that 1000 pF) such as those caused by 
long runs of shielded cable. For long cable runs, use a 4-20 mA 
output model.

The DPG1000DR and F4DR are designed for continuous operation. 
The gauge is powered on whenever a supply voltage is applied. 
Warm-up time is negligible. During normal operation the system 
pressure is displayed on the LCD and a corresponding output signal 
is present. 
DPG1000DRBL and F4DRBL display backlighting are on whenever 
the power is on. The display backlighting will not be apparent under 
bright lighting conditions.

The output is a continuous analog signal based on the transducer 
output rather than the display. The output is filtered to improve 
noise immunity and has a response time of about 50 msec. 
Characteristics of the output types are listed below. Values are 
factory set. Note that –15 psi, –100 kPa, etc. are not achievable, but 
prevents running out of range at whatever full vacuum is reached.

Electrical Connection

The DPG1000DR and F4DR series can be powered by any 9 to 

32 VDC or 8 to 24 VAC 50/60 Hz power source. An inexpensive 

unregulated low voltage source can be used. The magnitude of the 

supply voltage has negligible effect on the gauge calibration as long 

as it is within the stated voltage ranges. 
Do not allow the gauge supply voltage fall below 9 VDC or 8 VAC RMS. 

Operation below these values may cause erratic or erroneous readings 

or output. Models with 4-20 mA output power the current loop. Use a 

power source with sufficient voltage to operate the current loop.
Connection is made with the four conductor cable at the gauge 

rear. This cable accommodates both the gauge power supply and 

retransmission output. 
Connect power as shown below. When using low voltage AC 

power, either polarity may be used. Use the correct polarity with 

a DC supply.
Connect the retransmission output as shown below. Use of the 

shield (drain) wire is optional. It is not generally needed for 4-20 mA 

current loops unless very long cable lengths are used in electrically 

noisy environments.
The power supply (–) l ead is tied to the retransmission output 

ground. Therefore, if a DC supply is used, the power supply (–) lead 

should be considered common with regard to the retransmission 

output (–) connection.
If the analog output is not required, the transmitter will function as 

a low voltage powered gauge. Protect the output wires to prevent 

a short circuit.

Using the Retransmission Output

Operation

Test Function

1.  See rear label of gauge for range.
2.  Remove the covers on the Zero and Span controls on the front of 

the gauge. Zero calibration must be done before span calibration. 
If desired, gauges can be zeroed without doing span calibration. 

3. Connect gauge to a 9-32 VDC or 8-24 VAC 50/60 Hz power 

source. The magnitude of the supply voltage has negligible 
effect on the gauge calibration as long as it is within the stated 
voltage ranges. Over voltage may result in damage.

4. 

Zero for gauge reference pressure or vacuum gauges:

 With 

the gauge port open to atmosphere, adjust the Zero potenti-
ometer for a display indication of zero with the minus (–) sign 
occasionally flashing. Output should be 4.0 milliamps for mA 
output models or 0 volts for voltage output models.

 

Zero for absolute reference gauges:

 Apply full vacuum to the 

gauge. Adjust the Zero potentiometer for a display indication of 
zero with the minus (–) sign occasionally flashing. Output should 
be 4.0 milliamps for mA output models or 0 volts for 0-2 volt 
output models.

 

Zero for bipolar gauges:

 Apply full vacuum to the gauge. 

Adjust the Zero potentiometer for a display indication of full 
vacuum. Output should be 4.0 milliamps for mA output models  
or 0 volts for 0-2 volt output models.

5. 

Span for gauge reference pressure, absolute reference, and 
bipolar gauges:

 Apply full-scale pressure and adjust the Span 

potentiometer for a display indication equal to full-scale pres-
sure. Output should be 20.0 milliamps for mA output models or 
2 volts for voltage output models.

 

Span for gauge reference vacuum gauges:

 Apply full vacuum 

to the gauge. Adjust the Span potentiometer for a display indica-
tion equal to full-scale vacuum. Output should be 20.0 milliamps 
for mA output models or 2 volts for voltage output models.

6.  Verify pressure indications at 0%, 25%, 50%, 75%, and 100% 

of full scale and repeat calibration as needed to achieve best 
accuracy over desired operating range.

7.  Internal Zero and Span potentiometers adjust the analog out-

put to agree with display. These normally do not need to be 
adjusted. If the output does need adjustment, remove the rear 
cover to access the potentiometers. See image below. Adjust 
output Span and then Zero after gauge calibration is complete.

8.  Replace the potentiometer covers, rear cover and screws, taking 

care not to pinch the wires between the case and the rear cover.

When the front-panel TEST button is held depressed, the display 
and analog output are switched, independent of the system pres-
sure, to a test level determined by the setting of the Test potenti-
ometer. This test mode will allow setup and testing of the output 
by switching to this test level whenever desired without having to 
alter the system pressure.
To set the test output level, see gauge label for location of Test 
potentiometer. Remove the Test potentiometer cover. Press and 
hold the front-panel TEST button and adjust the Test potentiometer 
to set the display and output to the desired test level.

8 12 16 20 24 28 32

Supply Voltage (DC or AC RMS)

Voltage Compliance for 4-20 mA Retransmission Output

1400

1200

1000

800

600

400

200

0

DC Backlit & Non-Backlit

AC Backlit

AC

Output 

Span

Output 

Zero

Precautions

4

  Read these instructions before using the gauge. Configuration may 

be easier before installation. Contact the factory for assistance.

4

  These products do not contain user-serviceable parts. Contact us 

for repairs, service, or refurbishment.

4

  Gauges must be operated within specified ambient temperature ranges.

4

  Outdoor or wash down applications require a NEMA 4X gauge or 

installation in a NEMA 4X housing.

4

  Use a pressure or vacuum range appropriate for the application.

4

  Use fittings appropriate for the pressure range of the gauge. 

4

  Due to the hardness of 316 stainless steel, it is recommended that 

a thread sealant be used to ensure leak-free operation.

4

 For contaminated media use an appropriate screen or filter to 

keep debris out of gauge port. 

4

  Remove system pressures before removing or installing gauge.

4

  Install or remove gauge using a wrench on the hex fitting only. Do 

not attempt to turn gauge by forcing the housing. 

4

 Good design practice dictates that positive displacement liquid 

pumps include protection devices to prevent sensor damage from 
pressure spikes, acceleration head, and vacuum extremes.

Avoid permanent sensor damage! Do not apply vacuum to non-
vacuum gauges or hydraulic vacuum to any gauges. 

Avoid permanent sensor damage! NEVER insert objects into 
gauge port or blow out with compressed air.

 Gauges are not for oxygen service. Accidental rupture of sensor dia-

phragm may cause silicone oil inside sensor to react with oxygen.

p

 NEVER connect the gauge wires directly to 115 VAC or perma-

nent damage will result.

RED +

BLACK – 

GREEN –

WHITE +

Optional 

shield 

connection

Power (+) 

Power (–)

Remote Device

+

com.

Range Type

Output 

Option –15 psi

“0” on 

display

Full  

pressure

Gauge reference 
pressure

-I

n/a

4 mA

20 mA

-V

n/a

0 V

2 V

Gauge reference 
vacuum

-I

20 mA

4 mA

n/a

-V

2 V

0 V

n/a

Absolute reference

-I

4 mA

4 mA

20 mA

-V

0 V

0 V

2 V

Gauge reference 
vacuum/pressure

-I

4 mA

12 mA

20 mA

-V

0 V

1 V

2 V

Output Ranges

cecomp.com

1220 American Way Libertyville, IL 60048 

800-942-0315

CECOMP

 

Div. of

 

A

BSOLUTE

 P

ROCESS

 I

NSTRUMENTS

Cecomp maintains a constant effort to upgrade and improve its products. 

Specifications are subject to change without notice. See cecomp.com for 

latest product information. Consult factory for your specific requirements.

WARNING:

 This product can expose you to chemicals including 

lead, nickel and chromium, which are known to the State of 

California to cause cancer or birth defects or other reproductive 

harm. For more information go to www.P65Warnings.ca.gov

Reviews: