Part number 20220, Rev 0 
 

 

QUICK TROUBLESHOOTING GUIDE 

Check these items before calling for service 

PROBLEM: 

POSSIBLE CAUSE / SOULTIONS: 

Unit does not run 

 

Electrical circuit is not 110-120V 60Hz. 

 

The power cord is not plugged in. 

 

No power at electrical outlet. Check to make sure 
breaker is not tripped or fuse is not blown. Additionally, 
make sure unit is not plugged into a Ground Fault Circuit 
Interrupter (GFCI) type of outlet.  

Unit does not maintain at the 
proper temperature 

 

Check the room temperature.  We recommend the 
refrigerator or freezer should be placed in the air 
conditioned room between 65°F to 85°F.  If the room 
temp is too warm, the refrigerator or freezer may not be 
able to maintain the interior temp at proper range.   

 

Door is not closed properly. 

 

Amount of stored product is overloaded. 

 

Product replacements are pushed against rear wall or 
interrupted the proper refrigerator air circulation.  For 
the proper air circulation, place the products evenly on 
each shelf.  Do not push against the refrigerator’s rear or 
side walls. 

 

Evaporator is blocked by frost or ice.  Remove the 
products, unplug the refrigerator or freezer power, and 
allow the unit to defrost.  If the problem still exists, call 
for service.  

Appliance runs too long 

 

Prolong door openings. 

 

Control set too cold. 

 

Room temperature is high which will make the unit work 
harder to keep cool.  

Temperature of external wall 
surface is warm 

 

The exterior walls can be as much as 30 degrees warmer 
than room temperature due to the embedded 
condenser coils. This is normal when the unit is 
operating.   

Compressor noises 

 

Check the unit level. Also ensure none of the pipes in 
the compressor area are touching each other or against 
the side.  

 

Compressor may be overheated.  Please check the room 
temp and ensure the range is within 65°F to 85°F.   

 

A buzzing noise may indicate improper voltage to the 
compressor. Check supply voltage.  

 

If the problem still exists, call for service. 

Summary of Contents for PH-ABT-UCFS-0204G

Page 1: ...driver at time of delivery Make sure you get a signed copy Send a written request for an inspection to the carrier B Concealed Damage This type of damage may not be discovered until the unit is being...

Page 2: ...EPARTMENT please have the unit s model and serial number ready as well as the problem description The model and serial number is located on the serial tag which can be found on the interior left OR ri...

Page 3: ...y receipt acknowledging the damage This has to happen at the time of delivery or it won t happen at all Keep a copy for your records and send another to the carrier s damage claims department along wi...

Page 4: ...on this unit POWER SOURCE The supply circuit to this cabinet must conform to all National and Local Electrical Codes Consult the cabinet Serial Data plate for voltage cycle phase and amp requirements...

Page 5: ...product in the cabinet The motor compressor may start and stop several times when the unit is first started or after defrosting especially if the weather is very hot This is normal The motor compresso...

Page 6: ...perly The digital temperature controller is located on the front panel of the refrigerator It normally displays the current temperature inside of the unit and has several additional functions includin...

Page 7: ...ng 25 minutes will take the unit out of the defrost cycle PROGRAM MODE Note Most of the functions in Program Mode are for factory use only and should not be adjusted This procedure will show you how t...

Page 8: ...ides of the unit to allow for proper air flow and therefore more even cooling of the product When loading the top shelf avoid blocking the evaporator fan s if installed There should be at least 4 inch...

Page 9: ...the proper refrigerator air circulation For the proper air circulation place the products evenly on each shelf Do not push against the refrigerator s rear or side walls Evaporator is blocked by frost...

Page 10: ...bove 40C 104F Stop using immediately The display shows LL and the compressor does not run The ambient temperature is below 40C 40F Stop using immediately MOISTURE DURING THE SUMMER SEASON The amount o...

Page 11: ...ter out of sponge or cloth before cleaning around controls light bulb or any electrical parts Glass Doors if applicable Soap and water Clean the glass with a mild detergent and water on a soft cloth o...

Page 12: ...e paid by the customer ADDITIONAL THREE YEAR COMPRESSOR WARRANTY In addition to the two 2 year warranty stated above Horizon Scientific Inc warrants its hermetically and semi hermetically sealed compr...

Page 13: ...t damage during transit or installation fire flood or other external causes It does not apply to defects resulting from failure to properly install operate or maintain the product in accordance with t...

Reviews: