background image

 

DOOR BINS (not applicable to all models) 

Your unit may come equipped with storage bins in the door. Because of the temperature 

characteristics of these units, items stored in these bins may not be maintained in the desired 

temperature band. Please restrict the use of these bins for non-critical temperature storage. 

 

Shelf Installation (not applicable to all models) 

Hook shelf clips onto pilaster strips (see Illustration). Position all four shelf clips equal in 

distance from the floor for flat shelves. Wire shelves are oriented so that cross support bars are 

facing down. 

 

Place shelves on shelf clips making sure all corners are seated properly. 

 

 

Temperature control 

NOTE:  Temperature controls are preset at the factory to provide the desired air and product 
temperatures inside of the units and require no further adjustment. Please contact us before 
making any adjustments to determine if adjustment is necessary and, if so, to make sure it is 
performed properly.  

Both refrigerators and freezers have a manual thermostat located inside a box on the back of 
the unit. This thermostat controls the compressor. Turning the thermostat towards the higher 
numbers will make the unit colder. Lower numbers make it warmer. Turning the thermostat to 
the off or 0 position will turn the unit off. For dual-temp units (refrigerator and freezer), this 
thermostat controls the compressor based on freezer temperature. Temperature in the 
refrigeration compartment is controlled by venting air from the freezer compartment 

Summary of Contents for ABT-FC-30

Page 1: ...rrier s agent driver at time of delivery Make sure you get a signed copy Send a written request for an inspection to the carrier B Concealed Damage This type of damage may not be discovered until the...

Page 2: ...unit s model and serial number ready as well as the problem description The model and serial number is located on the serial tag which can be found on the interior left upper wall of the unit For con...

Page 3: ...STORING UNSEALED CONTAINERS OF CORROSIVE MATERIALS CAN CAUSE PERMANENT DAMAGE TO THIS REFRIGERATION UNIT AND WILL VOID THE WARRANTY ALWAYS WEAR SAFETY GLASSES WHEN WORKING WITH THIS EQUIPMENT ANY UNAU...

Page 4: ...knowledging the damage This has to happen at the time of delivery or it won t happen at all Keep a copy for your records and send another to the carrier s damage claims department along with a formal...

Page 5: ...SHOULD NOT VARY MORE THAN 5 FROM SERIAL PLATE RATINGS A separate circuit is recommended to prevent possible loss of product due to over loading or failure of other equipment on the same circuit PROTE...

Page 6: ...rature control NOTE Temperature controls are preset at the factory to provide the desired air and product temperatures inside of the units and require no further adjustment Please contact us before ma...

Page 7: ...ng include Leave about 2 3 inches of free space along the back and sides of the unit to allow for proper air flow and therefore more even cooling of the product When loading the top shelf avoid blocki...

Page 8: ...re pushed against rear wall or interrupted the proper refrigerator air circulation For the proper air circulation place the products evenly on each shelf Do not push against the refrigerator s rear or...

Page 9: ...oor will not close The unit is not level Refer to the Leveling section at the beginning of this manual Check for dirt and debris or items passing through the door seal MOISTURE DURING THE SUMMER SEASO...

Page 10: ...ovable shelves in dishwasher Exterior and Handles Soap and water Non Abrasive Glass Cleaner Do not use commercial household cleaners ammonia or alcohol to clean handles Use a soft cloth to clean smoot...

Page 11: ...or drain Replace the drain plug when defrosting is complete so warm air does not enter the freezer through the hole Defrosting tips If the frost is soft remove it with a plastic scraper If the frost i...

Page 12: ...n to the two 2 year warranty stated above Horizon Scientific Inc warrants its hermetically and semi hermetically sealed compressors to be free from defects in both material and workmanship under norma...

Page 13: ...external causes It does not apply to defects resulting from failure to properly install operate or maintain the product in accordance with the printed instructions provided or damage caused by the sto...