E-40
ElektroPhysik
14. Language selection
Three operator selectable languages are available:
-
German
-
English
-
French
Make your language selection as follows:
-
Switch off the instrument.
-
Press both ARROW-keys continuously while you
switch the instrument on. The current language
selection appears on display, “English” for
example.
-
Keep both ARROW-keys pressed down and
release keys as soon as your required language
is shown.
-
Your selection is valid without need to confirm.
15. Error messages
Error messages appear on the alphanumerical display.
Maximum
Calibration (CAL), sensitivity
(SENS), de-plating speed (SPEED)
or measuring area (AREA) have
reached maximum value
No cell
No measuring cell connected
No test current
No current flow in the solution. The
cause may be:
Clamp (38) is not connected to the
test sample.
Air bubbles on the surface.
No solution available in the
measuring cell (44).
Another solution being in the cell.
Surface insulated by grease, paint,
or other insulating material.
Cell voltage exceeding 10Volts.