ElektroPhysik
E-27
-
After this, use a caliper to measure diameter and
length of the de-pated wire area.
-
Use the measured wire dia. „d“ and the length „l“
of the de-plated wire area to determine the actual
deplated wire area as follows:
F = d x
πππππ
x l
F = Area
d = Diameter of wire
π
= 3,14
l = length of de-plated area
-
Use ARROW-keys for setting the actual
measuring area, e.g. 5,36 mm
2
.
-
Press SET-key to show the reading as calculated
by means of the actual measuring area, e.g. : 7,4
Insulating tape
Measure dia.
of
de-plated wire area
“d”
Measure length of
de-plate wire area
“l”
5,36mm²
Actual
measuring area
5,36
mm²
SET
7,4
Sn K 5,7 * * µm
Reading
Measuring area?