background image

 2 Wire Radar Sensors  
 User Instruction Manual 

Calibration — 4 -20 or 20 - 4 mA Output 

    

 FULL — Calibrate 20 mA or 4mA (Set Near Target)   
 1. Calibration mode LED color is blinking Green.  
      (for Radar Low Dielectric Materials has to be off)    
 2. Push button and  hold until LED turns Yellow (20 mA) 
     or push button and hold until LED turns  Red  (4 mA) 
 3. Release button, observe LED flashes to acknowledge  
     the calibration.       
 EMPTY— Calibrate 4 mA or 20 mA (Set Far Target) 
 1. Calibration mode LED color is blinking Green  
     (for Radar Low Dielectric Materials has to be off)    
 2. Push button and hold until LED turns   Red  (4 mA) 
     or push button and hold until LED turns Yellow (20 mA)  
 3. Release button, observe LED flashes to acknowledge 
     the calibration. 

   

For Radar to turn the Low Dielectric Materials operation mode         
  ON and OFF (this mode is recommended for materials with 
     dielectric constant lower than 4 and also to eliminate multiple  
     reflections in tank.)   

 

 1) To turn the Low Dielectric Materials ON. Push button and   
     hold until LED goes OFF after the sequence of Yellow ,Red   
     and turns Off. The Low Dielectric Material operation is On  
     when the LED’S Green light gives  two short blinks. 
 2) To turn the Low Dielectric Materials OFF. Push button and    
     hold until LED goes OFF after the sequence of Yellow , Red    
     and Turns OFF. The Low Dielectric Material operation is OFF  
     when LED is blinking Green. 
 3) Use “Hart” communication software.

         

730 The Kingsway Peterborough , Ont.  K9J6W6  Canada 
Tel: (705) 740 — 2010           Web: www.abmsensor.com 
Fax: (705) 740 — 2563       E-mail: [email protected]   

DWG 10A834R1 

Fig. # 1 - “2 Wire Sensor” Wiring  Connection 

Typical Installation

 

 

 

 

1)Radar unit must be installed into metal fitting with the antenna pointing downward.  

Power Input 
16 -  30 Vdc 
Conduit Entry 
 1/2” NPT  
Process Mtg. 
 1 1/2”,2”NPT 
Radar Rod  
 Antenna 

 

 

A
C
C
U
R
A
C











Per 
Chart 

Deadzone    

+/—0.25% 

EMPTY 
4 or 

20 or 4 mA 

FULL 

Level 

Material     

 
S
P
A

LED  

Re
d

 

Y

e

llo

 

Operation - electromagnetic pulse is transmitted from the ABM sensor . The pulse 
travels to the surface being monitored and is reflected off this surface back to the 
sensor . The time of flight is divided by 2  and converted to an output signal directly 
proportional to the material level .    

 

Calibration Pushbutton  

  Led Indicator For 4 - 20 mA  
  (750 ohm Max.)   

FCC INFORMATION TO RADAR USERS  
NOTE: This equipment has been tested and found to comply with the limits for a 
Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are 
designed to provided reasonable protection against harmful interference when 
the equipment is operated in a commercial environment. This equipment gener-
ates, uses, and can radiate radio frequency energy and, if not installed and used 
in accordance with the instruction manual, may cause harmful interference to 
radio communications. Operation of this equipment in a residential area is likely 
to cause harmful interference in which case the user will be required to correct 
the interference at his own expense. 

 
WARNING- Changes or Modifications not expressly approved  by  
ABM   Sensor Technology Inc. could void the user’s authority to 
operate the equipment.  

Wiring Information  
- Ground shield at one end only.  
- All terminal block wiring must be rated for 250V. 
- Terminal is for use only with equipment which has no live parts which   
are accessible . 
- Terminal is for use with equipment which maintains basic insulation 
from  hazardous voltage under normal and single fault conditions . 
- Connection used at the remote end of external  circuit . 
Recommended Wiring  
- 2 Wire shielded 24 AWG , 300 V 

 

Top View of Sensor (Access Cover Removed) 

Status 

Calibration Pushbutton  

 NOTE — Use only  
1/2” NPT  Conduit          

    

Terminal Block 

ABM 

SENSOR TECH. INC 

STATUS 

SWITCH 

CALIBRATION 

      +

 

        —  

 

4-20 mA  

V max. = 30 Vdc 
V min.  = 16 Vdc 

Radar 2 Wire 

Wire Belden 2 wire Shielded 9501 - 
Connect

 

Shield to Ground at one End 

 

R

  LOAD 

= Vsup. — 14(V) 

                    25 mA 

“2 Wire Radar” 

 

Loop Instrumentation 

 

+  

-- 

R Load 

Red 

Wht.  or Blk. 

 

+  

-- 

 

V supply   
16– 30 Vdc 

   HART    
 Interface 

RS232 to PC 

Reviews: