ABERLINK LTD
|
VATCH LANE
|
EASTCOMBE
|
GLOUCESTERSHIRE
|
GL6 7DY
|
UK
Tel : +44 (0)1453 884461
|
Fax : +44 (0)1453 882348
|
Email : [email protected]
|
Web : www.aberlink.com
Registered in England No. 7642245
|
VAT Registration No. GB 108 3075 37
The purpose of levelling the granite is to aid the dynamics of the machine, so the motors
are not working against gravity. High tolerance levels are not necessary. Check in y
direction and x direction, and use the nuts on the studding to adjust. Once done, fit the
last M10 nut and tighten against the underside of the bench.
Locate the regulator set and fit to the rear
of the machine in the holes provided.
See below for benches manufactured after
2014
Prior to pressure sensor
8mm O/dia tube to airbearings
–
connect pipe from the machine
6mm O/dia tube to the z
balance
– connect pipe
from the machine
Short length of
6mm tube
Air supply in.
Fitted with 8mm
push connector,
placed in ¼ bsp
port
Tee piece
Solenoid
block, if CNC
= Air Flow
Regulator with filter and
gauge
Continuous supply to z balance