Table 17. Guidance and manufacturer’s declaration –
proximity fields
Proximity Test IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment
Guidance
IEC 61000-4-3 385 MHz:
27 V/m @
18 Hz pulse
modulation
27 V/m
Recommended
minimum separation
distance d = 0.3 m
450 MHz:
28 V/m @ FM
modulation
28 V/m
710 MHz,
745 MHz,
780 MHz:
9 V/m @
217 Hz pulse
modulation
9 V/m
810 MHz,
870 MHz,
930 MHz:
28 V/m @
18 Hz pulse
modulation
28 V/m
1720 MHz,
1845 MHz,
1970 MHz:
28 V/m @
217 Hz pulse
modulation
28 V/m
2450 MHz:
28 V/m @
217 Hz pulse
modulation
28 V/m