Table 16.
Guidance and manufacturer’s declaration –
electromagnetic immunity
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment Guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
The device is intended
for use in the
electromagnetic
environment in which
radiated RF
disturbances are
controlled. The
customer or the user of
the device can help
prevent electromagnetic
interference by
maintaining a minimum
distance between
portable and mobile RF
communications
equipment
(transmitters) and the
device as
recommended below,
according to the
maximum output power
of the communications
equipment.
6 Vrms
(ISM/Radio
bands
between
150 kHz to
80 MHz)
80% AM at
1 kHz
6 Vrms
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to
2.7 GHz
80% AM at
1 kHz
10 V/m