Changing a Pain Management Program
d
Note: You cannot change a program when the pump is in
FULL
or
NEW CONTAINER
Lock.
With a
RATE CHANGE
Lock, the continuous delivery rate and bolus amount can be changed
within the set ranges.
Can be changed at any time
Cannot be changed
• Delivery rate
• Bolus
• Bolus lockout
• Container size
• Air sensitivity
• Dose limit amount
• Delivery mode
• Unit of measure
• Concentration
• Dose limit type
d
Note: Although the delivery mode cannot be changed, if a Cont Bolus therapy is
programmed, the continuous rate or bolus amount can be changed to zero. This function may
be restricted by the minimum values set with a
RATE CHANGE
Lock.
To change a program:
1
From the
STOP
mode, press [
CHANGE
] and select [
4
]
CHANGE PROGRAM
.
2
As each program parameter displays, make the desired changes, or press
[
YES/ENTER
] to accept the current setting.
3
Review the program.
4
Press [
START
] to begin delivery.
Pain Management
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
41
Summary of Contents for GEMSTAR
Page 7: ...vi Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Page 21: ...14 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Page 35: ...28 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Page 65: ...58 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Page 91: ...84 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Page 95: ...88 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Page 105: ...98 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Page 109: ...102 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Page 143: ...136 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Page 145: ...138 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...