7
If necessary, tap the cassette to eliminate any remaining
air bubbles.
8
Return the cassette to the upright position. Continue
priming until fluid fills the tubing.
9
Push down on the flow stop
to ensure the cassette
is in the closed position.
10
Insert the cassette into the cassette pocket. Refer to
“Loading the Cassette” on page 26.
CAUTION: When priming is complete, ensure no fluid flows at the distal end of
the pump set. If flow is observed, DO NOT use the pump set.
11
Connect the pump set to the patient-access device.
22
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
System Setup