FreeStyle Libre 3 Sensor - FCC ID: QXS-LIB03S
The System needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided in this manual.
Portable and mobile RF communications equipment can affect the System.
Use of accessories, transducers, and cables other than those specified or provided by Abbott Diabetes
Care could result in increased electromagnetic emissions or decreased electromagnetic immunity of the
System and result in improper operation.
The System should not be used adjacent to or stacked with other equipment and that if adjacent or
stacked use is necessary, the System should be observed to verify normal operation in the configuration
in which it will be used.
The device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) The device may not cause harmful interference, and (2) the device must accept any interference
received, including interference that may cause undesired operation.
Changes or modifications not approved by Abbott could void the user’s authority to operate the
equipment.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The System is intended for use in the electromagnetic environment specified below. The customer or the
user of the System should assure that it is used in such an environment.
Emissions test: RF emissions; CISPR 11
Compliance: Group 1
Electromagnetic environment – guidance: The System uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic
equipment.
Emissions test: RF emissions; CISPR 11
Compliance: Class B
Electromagnetic environment – guidance: The System is suitable for use in all establishments, including
domestic establishments and those directly connected to the public low voltage power supply network
that supplies buildings used for domestic purposes.
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The System is intended for use in the electromagnetic environment specified below. The customer or the
user of the System should assure that it is used in such an environment.
Immunity test: Electrostatic discharge (ESD); IEC 61000-4-2
IEC 60601 test level: ± 8 kV contact; ± 15 kV air