Service and Maintenance
Nonscheduled Maintenance Procedures
Section 9
Materials Required
1. A large container filled with approximately 500 mL of warm, deionized
water
2. DYN-A-WIPE lint-free pads (or other lint-free pads)
3. Distilled water
4. Small container of each reagent to refill the clean syringes
5. Phillips-head screwdriver (#2)
6. Flat-head screwdriver
Disabling the Analyzer
1. In the
MAIN MENU
, press
[SPECIAL PROTOCOLS]
followed by
>
DIS/ENAB
ANALYZER
@
and
[DISABLE ANALYZER]
2. To gain access to the syringe assemblies, open Left and Right Front Covers
as follows. Gently pull on cover using hand hold and swing door open about
30-45 degrees. Lift door fully upward from bottom area closest to door hinge
bracket, then rotate door fully open so lift and hold feature on upper door
hinge mechanism can be engaged. Lower door gently. Gently pull the Tower
Cover and lift the cover up and off the instrument. (Tower Door on the CS
model must be open.)
a. On the CS model, remove the four Phillips-head screws holding the Front
Skirt to the instrument and remove the skirt.
b. The skirt on the SL model has two removable sections adjacent to the
Flow Panel. Remove the section on the right side of the Flow Panel by
lifting it up and off its holding pins.
For easy access to the Flow Panel, the Sample Loader on the SL Model can
be pulled away from the analyzer by performing the following optional steps:
NOTE:
Instruments with serial number 59965AF and below do not have
the hardware configuration to support this option.
1. Remove the Phillips-head screw that holds the Sample Loader Tower
in place on the Front Panel. Refer to the following Figure 9.14.
Figure 9.14 Sample Loader Tower
2. Loosen the flat-head screws, two on each side of the instrument.