28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURES VP - 11.1 Barcode Check
3 - 144
3
6. Perform Barcode Alignment Test.
a. From Data Station Main Menu press:
-
*DIAGNOSTICS
-
*MORE
-
*MORE
-
*MORE
-
*MORE
-
*BARCODE ALIGNMENT
b. At Barcode Alignment Screen:
-
Type TEST (all capital letters).
-
Press [ENTER].
-
Press START key on Sample Loader.
When barcode read is successful, TEST is displayed.
-
Lightly shift rack left and right.
TEST must display every time.
“VP - 11.2 Barcode Adjustment” (Page 3-145)