2

Warnings

• This device was sterilized with ethylene oxide and is for single use only. Do not reuse or resterilize this device. Attempts to 

resterilize this device can cause a malfunction, insufficient sterilization, or harm to the patient.

• Do not use this device if the sterile package is open or damaged.
• Use the hemostasis valve to impede the backflow of blood during the implant procedure.
• Do not use a power injection syringe to inject contrast solution.

Precautions

• This device should be used only by physicians who are trained in standard transcatheter techniques. The physician should 

determine which patients are candidates for procedures that use this device. 

• The physician should exercise clinical judgement in situations that involve the use of anticoagulants or antiplatelet drugs 

before, during, and/or after the use of this delivery system. 

• Use on or before the last day of the expiration month that is printed on the product packaging label. 
• Use caution when advancing the sheath and dilator to avoid damaging tissue and vessels or interfering with previously 

implanted medical devices.

• Store in a dry place.

Potential Adverse Events

Potential adverse events that may occur during or after a procedure using this delivery system may include, but are not limited 
to:

Device Compatibility

The AMPLATZER

 devices compatible with the TorqVue

 LP delivery system are identified in Table 1.

CAUTION: No devices other than those listed in these tables have been tested for use with the delivery system. Using 
untested devices with the delivery system may result in technical failures and/or adverse events.

• Air embolism

• Infection

• Arrhythmia

• Myocardial infarction

• Arteriovenous fistulae

• Perforation

• Bleeding at the access site

• Peripheral pulse loss

• Brachial plexus injury

• Stroke/transient ischemic attack

• Cardiac tamponade

• Thrombosis

• Death

• Valve damage

• Dissection

• Vascular access site injury

• Endocarditis

• Vascular occlusion

• Hematoma

• Vessel damage

• Hemodynamic compromise

Table 1. Compatibility Chart for the TorqVue

 LP Delivery System and AMPLATZER

 Devices

Delivery System 

Sizes

4 Fr

5 Fr

AMPLATZER

 Vascular Plug

4 mm

6 mm

8 mm

AMPLATZER

 Duct Occluder II

03-04 mm

04-04 mm

03-06 mm

04-06 mm

05-04 mm

06-04 mm

05-06 mm

06-06 mm

Table 2. Delivery System Dimensions

Delivery system 

(catheter) size

Inner diameter of 

delivery catheter

4 Fr

1.17 mm (0.046 in)

5 Fr

1.50 mm (0.059 in)