4.

 

Use the valve bypass tool to open the hemostasis valve and advance the lead and stylet to the distal tip of the catheter. Use 
fluoroscopy to confirm the position of the lead within the catheter. 

5.

 

Aspirate for the presence of air and flush the catheter through the stopcock. Close the stopcock. 

Finding the Target Location 

1.

 

Connect the WorkMate Claris

 Recording System or the Merlin

 PSA to the catheter via the Supreme Electrophysiology Cable 

and adapter pins. 

2.

 

Use the deflection knob to deflect the distal tip of the catheter to the target location. 

3.

 

Adjust the position of the catheter until the target location is captured via the distal tip electrodes and confirm with pacing and 
EGM signals. 

Anchoring the Lead 

1.

 

After finding the target location, advance and anchor the lead helix into the target location. 

2.

 

Attach electrical connectors to the lead. 

3.

 

Confirm placement of the lead via EGM and pacing analysis. Use fluoroscopy to visually confirm location of the lead. 

4.

 

If repositioning is necessary, retract the lead helix and repeat earlier steps, as applicable, to reposition. 

5.

 

Once the lead is anchored and placement is confirmed, detach the electrical connections from the lead and catheter. 

 

1.

 

Lead Stabilization Channel: protects the lead and engages the shaft of the delivery sheath 

2.

 

Lead Retention Channel: passively maintains the lead position 

3.

 

Thumbpad: Actively maintains the lead position 

 

Removing the Catheter 

Table 4.  Slitting the Catheter 

 

1.

 

Partially retract the stylet to prevent lead dislodgement. 

 

2.

 

Provide slack to the lead by advancing the lead through the catheter. 

Summary of Contents for Agilis HisPro DS3H010-38

Page 1: ...Agilis HisPro Steerable Catheter With Electrodes Model DS3H010 38 Instructions for Use...

Page 2: ...e State of California to cause cancer and birth defects or other reproductive harm For more information go to www P65Warnings ca gov Indicates a trademark of the Abbott group of companies Indicates a...

Page 3: ...Pacing System Analyzer PSA Supreme Electrophysiology Cable 401980 401981 401982 or 401983 Adapter Pins Refer to the Merlin PSA Help Manual for a complete list of cables required to connect the PSA to...

Page 4: ...catheter may result in damage Handle with care Do not immerse the handle or cable connector in fluids electrical performance could be affected Do not expose to organic solvents Precautions to prevent...

Page 5: ...ckwise see figure below Do not use if the catheter is damaged 6 Turn the deflection knob counterclockwise to the neutral position Note After the catheter has been deflected the catheter shaft may not...

Page 6: ...ignals Anchoring the Lead 1 After finding the target location advance and anchor the lead helix into the target location 2 Attach electrical connectors to the lead 3 Confirm placement of the lead via...

Page 7: ...ction knob Secure the lead with thumb 6 Align the Slitter tool lead stabilization channel with the lead cup and secure the lead in the lead stabilization channel CAUTION Inaccurate alignment or exerti...

Page 8: ...e lead fully seated in the lead retention channel Lead not seated fully in the retention channel 9 Ensure the Slitter tool reference line is parallel with the catheter shaft see figure below Figure 2...

Page 9: ...f the Catheter shaft should have minimal curvature The Catheter should be slit using an axial motion avoiding rotation 11 Carefully remove the lead from the Slitter tool 12 Confirm placement of the le...

Page 10: ...symbols may be found on the product or product label For harmonized symbols refer to the Universal Symbols Glossary at https medical abbott manuals Symbol Description Outer dimension Inner dimension A...

Page 11: ...of use as set forth in the Product s Instructions for Use published by Abbott 4 The Product was used before the USE BEFORE date marked on the packaging of the Product Abbott will analyze the returned...

Page 12: ......

Page 13: ......

Page 14: ...St Jude Medical Cardiac Rhythm Management Division 15900 Valley View Court Sylmar CA 91342 USA 1 818 362 6822 2020 06 ARTEN600080436 B...

Reviews: