Agilis HisPro

 Steerable Catheter With Electrodes 

Instructions for Use 

Description 

The Agilis HisPro

 Steerable Catheter With Electrodes is a deflectable, slittable catheter with two distal tip electrodes. It serves as a 

delivery conduit for devices such as cardiac leads. The distal portion can be formed into a “U” shape, when fully deflected, to 
facilitate positioning of the tip at a desired location in the heart. The tip electrodes provide the ability to sense intracardiac 
electrograms (EGM) and pace when connected to the WorkMate Claris

 Recording System or the Merlin

 Pacing System Analyzer 

(PSA). The 10.5Fr catheter has a 7Fr inner diameter to allow delivery of a 6Fr cardiovascular transvenous device, including a 6Fr 
pacing lead. The catheter shaft is slittable allowing for removal following lead placement. 

Note: Safety and effectiveness of the Agilis HisPro catheter to deliver leads to the His Bundle has not been evaluated by the 
FDA. 

Table 1.  Agilis HisPro

 model 

Model 

Effective Length 

Outer Diameter 

Inner Diameter 

DS3H010-38 

38cm 

10.5Fr 

7Fr 

 
The catheter is compatible with the following Abbott Medical recording and pacing systems: 

Table 2.  Device Compatibility 

System Device 

Connect via 

WorkMate Claris

 Recording System 

Supreme Electrophysiology Cable (401980, 401981, 
401982, or 401983)  

Merlin

 Pacing System Analyzer (PSA) 

Supreme Electrophysiology Cable (401980, 401981, 
401982, or 401983), Adapter Pins 

 
Refer to the Merlin

 PSA Help Manual for a complete list of cables required to connect the PSA to the Adapter Pins. 

Indications for Use 

The Agilis HisPro

 Slittable Catheter With Electrodes is indicated to provide a pathway for delivery and support of transvenous 

devices within the chambers and vasculature of the heart and can be used for electrogram recording and stimulation. 

Contraindications 

Crossing a mechanical tricuspid valve with this device is contraindicated. 

Package Contents 

The following accessories are included with the Agilis HisPro

 Steerable Catheter With Electrodes. 

Table 3.  Package Contents 

Accessory 

Intended Use 

.035-inch guidewire 

Enables intravascular introduction of the dilator and Agilis HisPro Steerable Catheter With Electrodes. 

Dilator 

Dilates the vasculature and facilitates access to vasculature. 

Adaptor pins (2) 

Connects to the Merlin

 PSA via electrophysiology cable. 

Valve bypass tool (2) 

Aids entry of devices through the hemostasis valve. 

Slitter tool 

Slits the catheter for removal after delivery of the pacing lead. 

Stopcock 

Facilitates aspiration and flushing. 

Note: The guidewire material is PTFE coated type 304 stainless steel. 

Complications 

As with any catherization procedure, potential complications include thromboembolism, local and systemic infection, bleeding or 
hematoma at the puncture site, vascular dissection or perforation, cardiac perforation, and cardiac tamponade. 

Summary of Contents for Agilis HisPro DS3H010-38

Page 1: ...Agilis HisPro Steerable Catheter With Electrodes Model DS3H010 38 Instructions for Use...

Page 2: ...e State of California to cause cancer and birth defects or other reproductive harm For more information go to www P65Warnings ca gov Indicates a trademark of the Abbott group of companies Indicates a...

Page 3: ...Pacing System Analyzer PSA Supreme Electrophysiology Cable 401980 401981 401982 or 401983 Adapter Pins Refer to the Merlin PSA Help Manual for a complete list of cables required to connect the PSA to...

Page 4: ...catheter may result in damage Handle with care Do not immerse the handle or cable connector in fluids electrical performance could be affected Do not expose to organic solvents Precautions to prevent...

Page 5: ...ckwise see figure below Do not use if the catheter is damaged 6 Turn the deflection knob counterclockwise to the neutral position Note After the catheter has been deflected the catheter shaft may not...

Page 6: ...ignals Anchoring the Lead 1 After finding the target location advance and anchor the lead helix into the target location 2 Attach electrical connectors to the lead 3 Confirm placement of the lead via...

Page 7: ...ction knob Secure the lead with thumb 6 Align the Slitter tool lead stabilization channel with the lead cup and secure the lead in the lead stabilization channel CAUTION Inaccurate alignment or exerti...

Page 8: ...e lead fully seated in the lead retention channel Lead not seated fully in the retention channel 9 Ensure the Slitter tool reference line is parallel with the catheter shaft see figure below Figure 2...

Page 9: ...f the Catheter shaft should have minimal curvature The Catheter should be slit using an axial motion avoiding rotation 11 Carefully remove the lead from the Slitter tool 12 Confirm placement of the le...

Page 10: ...symbols may be found on the product or product label For harmonized symbols refer to the Universal Symbols Glossary at https medical abbott manuals Symbol Description Outer dimension Inner dimension A...

Page 11: ...of use as set forth in the Product s Instructions for Use published by Abbott 4 The Product was used before the USE BEFORE date marked on the packaging of the Product Abbott will analyze the returned...

Page 12: ......

Page 13: ......

Page 14: ...St Jude Medical Cardiac Rhythm Management Division 15900 Valley View Court Sylmar CA 91342 USA 1 818 362 6822 2020 06 ARTEN600080436 B...

Reviews: