background image

 

 

 

Storage and Handling 

Handle with care. Items should be stored at temperatures between -20ºC (-4ºF) and 50ºC (122ºF).  Store in a well-ventilated area 
under conditions that protect items from extreme humidity. In addition, cartons containing these items should be protected from 
water and should not be crushed. Do not remove from outer box. 

Packaging and Shelf-Life 

The catheter packaging is designed to prevent crushing of the product, to minimize product exposure to the atmosphere, and to 
provide for aseptic product transfer. It is recommended that the product remain in the unopened package until time of use. Contents 
are sterile if the package is unopened and undamaged. Do not resterilize. The expiration date is marked on the outside of the 
package. Dispose of used product and packaging following standard solid biohazard waste procedures. 

Instructions for Use 

Preparing the Catheter 

1.

 

Review the label carefully for product identification. Do not use if the device is expired. 

Note: International formatting has been used for the expiration date (YEAR-MONTH-DAY) 

2.

 

Inspect the package prior to use. Do not use if the package is opened or damaged. 

3.

 

Remove the components of the sterile package within the sterile field. 

4.

 

Inspect the catheter and both distal tip electrodes for integrity and overall condition. Do not use if the catheter is damaged. 

5.

 

Verify that the catheter deflects properly by rotating the deflection knob clockwise (see figure below). Do not use if the catheter 
is damaged. 

6.

 

Turn the deflection knob counterclockwise to the neutral position.  

Note: After the catheter has been deflected, the catheter shaft may not return to a fully straight position. 

7.

 

Connect the three-way stopcock to the luer fitting prior to flushing the catheter (Figure 1). Luer fitting is specifically for fluid, do 
not use luer port for guide wire delivery. 

CAUTION: Catheter should be flushed with the stopcock attached. 

8.

 

Flush the catheter with normal saline or heparinized saline via the stopcock. Close the stopcock. 

9.

 

Flush the dilator with normal saline or heparinized saline. 

10.

 

Insert the dilator into the hemostasis valve and advance the dilator until it extends beyond the distal tip of the catheter shaft. 

Figure 1.  Deflecting and Flushing the Catheter 

 

 

Note: Connect the three-way stopcock to the luer fitting prior to flushing the catheter. 

Positioning the Distal Tip 

1.

 

After gaining vascular access, load the dilator into the catheter using the hemostasis seal and advance the catheter and dilator 
over the guidewire to track the position of the guidewire, catheter, and dilator. 

CAUTION: Deflecting the catheter while the dilator is inserted may cause catheter damage. 

CAUTION: The dilator is designed to be loaded into the Agilis HisPro and advanced over the guidewire. 

2.

 

After reaching the right atrium, retract the dilator and guidewire from the catheter. 

CAUTION: Excessive manipulation of the catheter while retracting the dilator may lead to catheter damage. 

3.

 

Aspirate for the presence of air and flush the catheter through the stopcock. Close the stopcock. 

Summary of Contents for Agilis HisPro DS3H010-38

Page 1: ...Agilis HisPro Steerable Catheter With Electrodes Model DS3H010 38 Instructions for Use...

Page 2: ...e State of California to cause cancer and birth defects or other reproductive harm For more information go to www P65Warnings ca gov Indicates a trademark of the Abbott group of companies Indicates a...

Page 3: ...Pacing System Analyzer PSA Supreme Electrophysiology Cable 401980 401981 401982 or 401983 Adapter Pins Refer to the Merlin PSA Help Manual for a complete list of cables required to connect the PSA to...

Page 4: ...catheter may result in damage Handle with care Do not immerse the handle or cable connector in fluids electrical performance could be affected Do not expose to organic solvents Precautions to prevent...

Page 5: ...ckwise see figure below Do not use if the catheter is damaged 6 Turn the deflection knob counterclockwise to the neutral position Note After the catheter has been deflected the catheter shaft may not...

Page 6: ...ignals Anchoring the Lead 1 After finding the target location advance and anchor the lead helix into the target location 2 Attach electrical connectors to the lead 3 Confirm placement of the lead via...

Page 7: ...ction knob Secure the lead with thumb 6 Align the Slitter tool lead stabilization channel with the lead cup and secure the lead in the lead stabilization channel CAUTION Inaccurate alignment or exerti...

Page 8: ...e lead fully seated in the lead retention channel Lead not seated fully in the retention channel 9 Ensure the Slitter tool reference line is parallel with the catheter shaft see figure below Figure 2...

Page 9: ...f the Catheter shaft should have minimal curvature The Catheter should be slit using an axial motion avoiding rotation 11 Carefully remove the lead from the Slitter tool 12 Confirm placement of the le...

Page 10: ...symbols may be found on the product or product label For harmonized symbols refer to the Universal Symbols Glossary at https medical abbott manuals Symbol Description Outer dimension Inner dimension A...

Page 11: ...of use as set forth in the Product s Instructions for Use published by Abbott 4 The Product was used before the USE BEFORE date marked on the packaging of the Product Abbott will analyze the returned...

Page 12: ......

Page 13: ......

Page 14: ...St Jude Medical Cardiac Rhythm Management Division 15900 Valley View Court Sylmar CA 91342 USA 1 818 362 6822 2020 06 ARTEN600080436 B...

Reviews: