PROGRAMMING…
Oxygen Calibration Sequence
Cell Constant mV
The upper display shows the millivolt offset of the oxygen probe from the last
successful calibration.
Span % of Theory
A value between 90 and 110% should be displayed. The display indicates the
oxygen probe output slope using parameters derived at the last successfull
two-point calibration or using the preset values.
Calibration User Code
[00000 to 19999]
If an incorrect value is entered, access to the calibration page is inhibited and
the display returns to the top of the oxygen calibration page.
One Point Calibration
Select the one-point calibration sequence.
Connect to Air
Connect the air supply to the probe (refer to the probe operating instructions,
IM/ZFG2). The upper display indicates the measured % oxygen.
Calibrating Air
The upper display indicates the measured % oxygen. When a stable reading is
detected the display advances automatically to the next frame. To abort
calibration, press either the
or
switch.
Calibration Pass/Fail
On completion a calibration status message is displayed:
Calibration Pass
Calibration sequence successful
Failed Constant
Cell offset >
±
10mV
(upper display shows cell mV output)
Failed Unstable
Cell output unstable (drifting).
Note. If the sensor calibration is unsuccessful then the 'Cell Constant mV' and
'Span % of Theory' parameters are unaffacted. The instrument continues to
operate using parameters stored during the last successful calibration.
Press
to return to the top of the
Oxygen Calibration Page.
Press
to advance to the
Secure Parameters Page.
Note. Before commencing a gas calibration procedure the analyzer and probe must be switched on and allowed to operate
for at least one hour to allow the system to stabilize thermally.
7.1
Single-point Calibration
The calibration sequence involves standardizing the analyzer and the oxygen probe, using air as the test gas. Until a calibration
sequence has been completed successfully, the existing slope remains unaffected.
Failed Unstable
Preset Cal.
Connect to Air
20.95
Calibrating Air
One Point Cal.
-----
Two Point Cal.
20.95
-----
Calibration Pass
Failed Constant
0000.0
Cal. User Code
-----
OXYGEN CAL. SEQ.
0.0
Cell Constant mV
100.0
Span % of Theory
SECURITY CODE