23
YES
xxxxx
mV Span 1 180mV
-----
ELECTRICAL CAL.
xxxxx
Calibrate NO
xxxxx
mV Zero 1 -20mV
Operating Page
xxxxx
mV Span 2 40mV
xxxxx
mV Zero 2 10mV
-----
Adjust RTX Zero
25.0
Adjust CJ Value
-----
Adjust RTX Span
8
CALIBRATION
8.3
Electrical Calibration Page
In this section the actual values denoted by
xxxxx
are not important and should only be used to determine display reading
stability.
Electrical Calibration Page
Calibrate
Select
YES
to access the electrical calibration sequence. Select
NO
to
advance to the
Adjust RTX Zero
parameter below.
mV Zero 1
Set the Cell simulator millivolt source to –20mV and allow the display to
stabilize.
mV Span 1
Set the Cell simulator millivolt source to 180mV and allow the display to
stabilize.
mV Zero 2
Set the Temperature simulator millivolt source to 10mV and allow the display to
stabilize.
mV Span 2
Set the Temperature simulator millivolt source to 40mV and allow the display to
stabilize.
Adjust Cold Junction Value
Set the measured ambient temperature in
°
C
Adjust RTX Zero
Set the milliammeter reading to the minimum retransmission level determined by
the
Retransmission Type
setting – see Section 7.6. For example, when the
Retransmission Type
is set to
4
–
20
, adjust the milliammeter reading to 4mA.
Adjust RTX Span
Set the milliammeter reading to the maximum retransmission level determined
by the
Retransmission Type
setting – see Section 7.6. For example, when the
Retransmission Type
is set to
4
–
20
, adjust the milliammeter reading to 20mA.
Press
to return to the top of the
Electrical Calibration Page.
Press
to advance to the
Operating Page.