Page 1
–18
2104349 rev. AD
The unit is designed to provide EMI/RFI protection of the low level signals and to
protect the circuitry from other environmental effects.
Figure 1
–15 Explosion proof multivariable transducer (EXIMV)
Table 1
–6 Explosion proof EXIMV specifications
Temperature limits
Compensated:
-20°F to 140°F (-29°C to 60°C)
Operational:
-40°F to 140°F (-40°C to 60°C)
Storage:
-40°F to 185°F (-40°C to 85°C)
Resolution
24 Bit A/D conversion (0.000012% FS) (0.0012% FS effective signal resolution)
Mounting spec.
Change from perpendicular (front to back around X-axis)
≤0.5% URL
Static pressure
Accuracy
Linearity, Hysteresis and repeatability combined: ± 0.075% of user-calibrated spans from
20% to 100% of URL
Ambient temperature
effect within the
operational
temperature limit
± 0.075% of URL +0.06% of Reading
Stability
± 0.1% of URL (for 12 months)
Differential pressure
Accuracy
Linearity, Hysteresis and repeatability combined: ± 0.075% of user-calibrated spans from
20% to 100% of URL.
Ambient temperature
effect within the
operational
temperature limit
± 0.075% of URL + 0.06% of Reading
Stability
± 0.1% of URL (for 12 months)