2012981-001 rev. AB
13
Figure 12 Setup Tab
8)
Move back to the Current Values tab within the Flow Measurement application,
and note the DP, SP and Tf values. These values should be the same as the
ones that the user just established.
9)
Next, move to the Commands tab within the Setup tab on the Flow
Measurement application, and perform a reset volume. This is accomplished by
setting the parameter to Yes from the drop-down selection and clicking the
Send button.
10)
Return to the Current Value tab. The flow rate will now reflect the flow based on
the values that were configured earlier under the Constants tab and the DP, SP
and Tf fixed values that were entered in the Holding Registers tab.
11)
Verify the flow rate against known values or results of a third party calculation.
12)
After verification is completed, return to the Calibration menu under Local
Connect, and change the Register entries back to their original values. For the
microFLO and XFC these will be:
•
SP – 7.4.1
•
DP – 7.4.0
•
Tf – 7.4.2