2012981-001 rev. AB
11
If the user chooses to move to the Events Collection view, they will be able to see the
Dp calibration check event, as entered.
Additionally, the user can use the drop-down menu to select Dp, Sp, Tf or DP & Sp.
By entering these values, the user will log the values into Events the event log.
4.2 Calibration
Tab
The Calibration tab (see Figure 10) allows the current measured values to be
calibrated and logged as events when they are compared to a known, benchmark
value.
Figure 10 Calibration Screen
Do not use Gauge selection for Sp found on the Checks and
Calibration tabs with liquid application as of PCCU 7.12.3.3. This
option is awaiting modification. Only Absolute should be used
until the PCCU software has been modified.
5.0 VERIFICATION
5.1 Verification
Tools
These hardware and software tools are required for testing the Liquid Orifice Tube
Measurement applications on the microFLO device.
•
microFLO 6213 boards (Part Number: 2100767-001*)
•
microFLO 6213 flash (Part Number: 2103938-001*)
•
Standard display – 2X24 LCD (Part Number: 2102804-004)
•
Hardware necessary to stimulate remote sense input
•
Multimeter for reading contact out states
•
PCCU32
(Current
build)