VortexMaster FSV430, FSV450 SwirlMaster FSS430, FSS450
| CI/FSV/FSS/430/450-EN Rev. G 43
Configuration of the analog- / HART-input
Easy Setup
Easy Setup
HART In Value
Analog In Value
Ext. T
Ext. T
Next
Edit Next
Edit
1. Use
to call up the edit mode.
2. Use
/
to select the desired function for the
analog-/ HART input.
HART In Value
Analog In Value
Function
Ext. T
Ext. T
External temperature
transmitter downstream for
energy measurement
Pressure Pressure
External
pressure
transmitter
Gas Content
Gas Content
External gas analyzer
Density
Density
External density transmitter
Int.T Int.T
External
temperature
transmitter upstream for
energy measurement
—
Ext. Cutoff
External output zero return
3. Confirm the selection with
.
In the following menus, the measurement range limits for the
external transmitters are fixed at the analog input.
Easy Setup
Upper value = 20 mA
Lower value = 4 mA
T Ext. Upper Range
xx.xx
Next
Edit
4. Use
to call up the edit mode.
5. Use
/
/
to set the measuring range limits for
the respective process value.
6. Confirm the selection with
.
Configuration of the parameters dependent on the
operating mode
The parameters shown in this position in the menu depend on
the selected operating mode and are not presented in detail
here. Follow the chapters "Operating mode" on page 47 and
Parameter descriptions in the operating instruction for detailed
information!
Select the end value for the current output
Not for devices with Modbus-communication!
Setting of the flow rate or energy quantity at which the current
output is to output 20 mA (100 %). The value entered must be
at least 15 % of Q...maxDN.
Easy Setup
Qvmax
5.00 l/s
Next
Edit
1. Use
to call up the edit mode.
2. Use
/
/
to set the desired end value for the
current output.
3. Confirm the selection with
.
Setting the damping value
Adjustment of the damping for the respective process value.
(the value relates to 1 T (Tau)).
The damping relates to a step change in the flow rate or
energy quantity or temperature.
The damping affects the instantaneous value in the process
display and at the current output.
Easy Setup
Damping Qv
1.0000 sec
Next
Edit
1. Use
to call up the edit mode.
2. Use
/
/
to set the desired damping for the
respective process value.
3. Confirm the selection with
.