Measuring channel parameters 11H...16H
Parameter address (Offset)
Data type
Function
Value range and coding
0000H
Byte
Input type
0H = off
01H = 0...20 mA
02H = 4...20 mA
03H =
±
2.5 mA
04H =
±
5 mA
05H =
±
20 mA
06H = 0...25 mV
07H =
±
25 mV
08H = 0...100 mV
09H =
±
100 mV
0AH = 0...500 mV
0BH = 0...2.5 V
0CH =
±
2.5 V
0DH = 0...5 V
0EH =
±
5 V
0FH =
±
10 V
10H =
±
20 V
11H = Pt100 I (−50..+150
°
C)
12H = Pt100 II (−50..+500
°
C)
13H = Pt100 III (−200..+850
°
C)
14H = TC B
15H = TC E
16H = TC J
17H = TC K
18H = TC L
19H = TC N
1AH = TC R
1BH = TC S
1CH = TC T
1DH = TC U
1EH = RS 485
0001H
Byte
Temperature unit
00H =
°
C
01H =
°
F
0002H
Byte
Physical unit of measurement
00H = free unit
01H = mA
02H = A
03H = mV
04H = V
05H = mbar
06H = bar
07H = Pa
08H = kPa
09H =
°
C
0AH =
°
F
0BH = K
0CH = l/s
0DH = l/min
0EH = %
0FH = ‰
10H = kW
11H = MW
12H = 1/min
13H = m
3
/h
0003H
Byte
Display format
00H = linear
01H = linear with 2 increments
02H = linear with 3 increments
03H = logarithmic
0004H
Byte
Enable channel display
00H = off
01H = on
0005H
Float
Measuring range start
-999...+9999
0009H
Float
Measuring range end
-999...+9999
Parameters
13