Start-Up Procedure
3
2. UNPACKING AND SETUP
a. VISUAL INSPECTION
The MR spectrometer system and its accessories may come in up to 4 different carrying cases depending on the
options selected at purchase.
CASE IDENTIFICATION
CONTENT
SPECTRORADIOMETER
MR100 OR MR200 SERIES SPECTRORADIOMETER
COMPUTER
COMPUTER
MONITOR
KEYBOARD
MOUSE
BACKUP SOFTWARE
COMPUTER MANUAL FROM ORIGINAL MANUFACTURER
DOS/WINDOWS AND BOMEM GRAMS SOFTWARE
MANUALS
ACCESSORIES
WIDE ANGLE TELESCOPE
MEDIUM ANGLE TELESCOPE
NARROW ANGLE TELESCOPE
MIR DETECTOR(MCT)
NIR DETECTOR(InSb)
CCD CAMERA
CCD CAMERA MONITOR
CCD CAMERA POWER SUPPLY
COLD SOURCE
DATA CABLE
WHITE LIGHT ALIGNMENT SOURCE & POWER SUPPLY
FUNNEL FOR FILLING LN
2
DETECTOR DEWARS
THE MANUAL YOU ARE PRESENTLY READING
TRIPOD
TRIPOD
Upon reception of the material check for any external damage to the carrying cases. If any damage is observed,
please notify the courier and have this stated on the packing slip. This is important for making a claim should the
equipment have been damaged during shipment. Please notify the After Sales Service Department at ABB.
Before unpacking the system a minimum delay of 12 hours is recommended to allow the equipment to
reach thermal equilibrium with the environment. This will help prevent any moisture condensation on the
optics. Most MR systems use ZnSe optics which are non hygroscopic but they can optionally be supplied with
KBr or other hygroscopic materials. When this is the case special precautions need to be taken to avoid
damaging the optics, please refer to “Handling and Care of Hygroscopic Materials”.