ABB
TB82PH ATEX Product Instruction Manual
TB82PH-ATEX-EN-A
13
CONFIGURE MODE
Before the instrument can be put into a process measurement loop it must be configured. The TB82PH Transmitter
operation (non-fieldbus versions) can be set in either a BASIC or ADVANCED Mode. BASIC mode is sufficient for 95% of
all measurements. ADVANCED functions are described in
Italic
print in the ‘alternatives’ section of the default
configuration table and in the ‘Configure Mode - Flow Chart for ADVANCED Mode’ section.
Figures 15
through
25
show
typical screen flows for a BASIC pH Configuration.
Default Configuration
Parameter
Default
Alternative
Mode
BASIC ADVANCED
Analyzer
pH
ORP, pIon,
Ion Concentration
Analyzer input
pH Glass
Antimony,
Custom sensor
Reference Impedance
Temperature Sensor
None
3K Balco
Pt100 RTD
Temperature
Compensation Type
Manual Auto,
Auto Solution
Output
4 mA = 0.00 pH
20 mA = 14.00 pH
Any 1 pH to 16 pH span,
any span above 100 mV from
–1999 mV to +1999 mV,
non-linear function generator
,
Damping
00.5 seconds
Adjustable to 99.9 seconds
Diagnostics
Off On
Safe Mode*
Fail low
Fail high
Spike*
0% Spike Magnitude
0 to 100% magnitude
Note: Alternatives in
Italics
are available in ADVANCED mode only
* Safe Mode and Spike functions not available on fieldbus versions
7.00
CONFIG
exit to
MEASURE
SELECT
MENU
pH
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
7.00
MODIFY
exit to
MEASURE
SELECT
NEXT
pH
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
7.00
Basic
exit to
MEASURE
SELECT
NEXT
pH
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
7.00
ANALZR
exit to
MEASURE
SELECT
NEXT
pH
Figure 15
Figure 16
Figure 17
Figure 18