FEP630, FEH630
ELECTROMAGNETIC FLOWMETER | OI/FEP630/FEH630-EN REV. D
93
Menu / parameter
Description
Input/Output / ...Dig.Out V1/V2 / ...Setup Logic Output
Input/Output / ...Dig. Out V3/V4 / ...Setup Logic Output
Logic Output Action
Selection of binary output function.
See description ‘„Input/Output / ...Dig.Out 41/42 / ...Setup Logic Output’.
Active Mode
Select switching properties for the binary output.
• Active High: Normally open
• Active Low: Normally closed
Default setting: Active High.
Input/Output / ...Dig.Out V1/V2 / ...Alarm Config
Input/Output / ...Dig. Out V3/V4 / ...Alarm Config
General Alarm
Select error messages signaled via the binary output V1 / V2 or V3 / V4.
Only if the parameter ‘Logic Output Action’ is set to Alarm Signal.
Qv Volumeflow Max
Qv Volumeflow Min
EPD
TFE
Gas Bubble
Conductivity
Sensor Temperature
In house Temp
Input/Output / ...Dig.In V1/V2
Input/Output / ...Dig.In V3/V4
Function
Select a function for the digital input.
• No function: No function.
• Reset all Totalizer: Counter reset for all counters (forward flow, reverse flow and difference totalizer)
• Stop
all
Totalizer
:
External counter stop for all counters (forward flow, reverse flow and difference totalizer)
• Auto. Zero Adjust: Start external zero point balancing.
• Set Flowrate to zero: Sets flow measurement to 0.
• Start/Stop Batching: Start / stop fill operation (only when FillMass function is activated).
• Dual Range Mass: Switchover Qm Max 1 / Qm Max 2.
• Dual Range Volume: Switchover Qv Max 1 / Qv Max 2.
Active Mode
Select switching properties for the digital input.
Delay Time
Selection of delay time for suppressing EMC faults on the digital input.