ABB FAM3200 Series Manual Download Page 4

FAM3200

 ARMORED VARIABLE AREA PURGEMETER  |  CI/FAM3200-EN REV. D 

 

 

… 1  Safety 

Intended use 

 

This device is intended for the following uses: 

•  For conveying liquids, gases (including unstable liquids 

and gases) and vapors. 

•  For flow measurement of the operating volume under 

constant operating conditions (pressure, temperature, 
density). An output of the flow is also possible in 
standard or mass units. 

 

 

The device has been designed for use exclusively within the 
technical limit values indicated on the name plate and in the data 
sheets. 
The following technical limit values must be observed: 

•  The permissible pressure (PS) in the permissible 

measuring medium temperature (TS) may not up-scale 
the pressure-temperature values (p/T ratings). 

•  The maximum or minimum operating temperature must 

not be up-scaled or down-scaled. 

•  The maximum ambient temperature must not be 

exceeded. 

 

 

When using measuring media, the following points must be 
observed: 

•  Measuring media may only be used if, based on the state 

of the art or the operating experience of the user, it can 
be assured that the chemical and physical properties 
necessary for operational security of the materials of the 
wetted parts of the flowmeter sensor will not be 
adversely affected during the operating time. 

•  Media containing chloride in particular can cause 

corrosion damage to stainless steels which, although not 
visible externally, can damage wetted parts beyond repair 
and lead to the measuring medium escaping. It is the 
operator's responsibility to check the suitability of these 
materials for the respective application. 

•  Measuring media with unknown properties or abrasive 

measuring media may only be used if the operator is able 
to perform regular and suitable tests to ensure the safe 
condition of the device 

 

 

The operator bears sole responsibility for the use of the devices 
in relation to suitability, intended use and corrosion resistance 
of the materials in relation to the measuring medium. 

 

The manufacturer is not liable for damage arising from improper 
or non-intended use. 
 

  

Repairs, alterations, and enhancements, or the installation of 
replacement parts, are only permissible insofar as these are 
described in this manual. Approval by ABB Automation Products 
GmbH must be sought for any activities beyond this scope. 
Repairs performed by ABB-authorized specialist shops are 
excluded from this. 
 

 

Improper use 

 

The following are considered to be instances of especially 
improper use of the device: 

•  Operation as a flexible compensating adapter in piping, 

for example for compensating pipe offsets, pipe 
vibrations, pipe expansions, etc. 

•  For use as a climbing aid, for example for mounting 

purposes. 

•  For use as a bracket for external loads, for example as a 

support for piping, etc. 

•  Material application, for example by painting over the 

housing, name plate or welding/soldering on parts. 

•  Material removal, for example by spot drilling the 

housing. 

 

 

Warranty provisions 

 

Using the device in a manner that does not fall within the scope 
of its intended use, disregarding this manual, using 
underqualified personnel, or making unauthorized alterations 
releases the manufacturer from liability for any resulting 
damage. This renders the manufacturer's warranty null and void. 
 

 

Manufacturer’s address 

 

ABB Automation Products GmbH 
Measurement & Analytics 

Schillerstr. 72 
32425 Minden 
Germany 
Tel:   +49 571 830-0 
Fax:  +49 571 830-1806 

 

 

Customer service center 

Tel:  +49 180 5 222 580 
Email:  [email protected] 

 
 

  

 

Summary of Contents for FAM3200 Series

Page 1: ...le Area Purgemeter Measurement made easy FAM3200 Introduction Armored variable area purgemeters for measuring small flows of gases and liquids Additional Information Additional documentation on FAM320...

Page 2: ...s 8 Pressure loss 8 Prevention of compression oscillations when measuring gases 8 Pressure shocks 8 Solids content in the measuring medium 9 Mounting 9 General information 9 Flowmeter installation 9 6...

Page 3: ...he product may only be performed if expressly permitted by these instructions Information and symbols on the product must be observed These may not be removed and must be fully legible at all times Th...

Page 4: ...measuring media may only be used if the operator is able to perform regular and suitable tests to ensure the safe condition of the device The operator bears sole responsibility for the use of the devi...

Page 5: ...follower system 6 Cable entry Figure 1 Setup example The devices are used for local indication of the current flow rate with integrated alarm signalling units as flow rate monitors or with differenti...

Page 6: ...he shipper without delay and before installation Transport The center of gravity of some devices is not at the center of the equipment Use any of the available attachment points on the device for tran...

Page 7: ...and sudden pressure surges Use valves which open slowly If the flowmeter is installed in a pipeline where decommissioning is impossible or inexpedient a bypass line should be provided For gaseous mea...

Page 8: ...ntion of compression oscillations when measuring gases During low flow amounts and low operating pressure so called compression oscillations of the float can occur If the maximum upstream pressure lis...

Page 9: ...installation Install the flowmeter centrally in the piping at the required position using the correct threaded connections When tightening the process connections on the flowmeter use a spanner of the...

Page 10: ...position transmitter in two wire technology Other cable lengths available on request 7 Commissioning CAUTION Risk of burns due to hot measuring media The device surface temperature may exceed 70 C 158...

Page 11: ...s Repair and maintenance activities may only be performed by authorized customer service personnel When replacing or repairing individual components use original spare parts Note Spare parts can be or...

Page 12: ...s conditions The process conditions for example high pressures and temperatures toxic and aggressive measuring media can give rise to hazards when dismantling the device If necessary wear suited perso...

Page 13: ...Material load Figure 7 PS pressure TS temperature Maximum permissible operating pressure The permissible operating pressure is dependent on the optional device equipment Model Option Max permissible o...

Page 14: ...0 7 10 15 1 0 14 5 8 0 12 04H 6 1 59 230 8 12 18 18 0 7 10 15 1 0 14 5 8 0 12 05H 10 2 64 350 12 36 18 18 0 7 10 15 1 0 14 5 8 0 12 06H 16 4 23 540 19 07 18 18 0 7 10 15 1 0 14 5 8 0 12 07H 25 6 6 850...

Page 15: ...4 0 58 02 0 3 4 35 85 1 23 31H 600 158 5 18000 635 66 8 8 4 0 58 02 0 3 4 35 95 1 38 32H 700 184 92 21000 5547 61 8 8 4 0 58 02 0 3 4 35 105 1 53 33H 800 211 34 24000 847 55 8 8 4 0 58 02 0 3 4 35 13...

Page 16: ...mentation declarations of conformity and certificates are available in ABB s download area www abb com flow Trademarks Buna N is a registered trademark of DuPont Dow Elastomers Kalrez and Kalrez Spect...

Page 17: ...Company Address Contact person Telephone Fax Email Device details Type Serial no Reason for the return description of the defect Was this device used in conjunction with substances which pose a threa...

Page 18: ...18 FAM3200 ARMORED VARIABLE AREA PURGEMETER CI FAM3200 EN REV D Notes...

Page 19: ...FAM3200 ARMORED VARIABLE AREA PURGEMETER CI FAM3200 EN REV D 19 Notes...

Page 20: ...ice With regard to purchase orders the agreed particulars shall prevail ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document We reser...

Reviews: