8.5  Replacing sensor assembly flow disk

Referring to Fig. 8.7:
1.

Place the sensor assembly on a clean flat surface with the 
flange studs pointing down. 

2.

Remove 2 long screws 

B

 using a small flat-headed 

screwdriver.

3.

Remove the flow disk assembly from flange manifold 

C

.

4.

Remove 2 (short) screws 

A

 using a small crosshead 

screwdriver.

5.

Separate flow disk cover plate 

D

, flow disk top cover 

gasket 

E

, flow disk 

F

, flow disk plate 

G

 and flow disk 

lower cover gasket 

H

. Discard gaskets 

E

 and 

H

.

6.

Fit a new flow disk into flow disk plate 

G

, aligning the 

center of the curved flow disk face 

I

 against the small 

hole in the middle of the plate and in the orientation shown 
(concave face uppermost).

7.

Place a new flow disk top gasket over flow disk plate 

G

.

8.

Apply a light coating of anti-seize compound (suitable for 
temperatures up to 200 °C [392 °F]) to the threads of 2 
(short) screws 

A

 and secure items 

D

E

F

 and 

G

  

using 2 (short) screws 

A

. Tighten the screws firmly.

9.

Place a new flow disk lower cover gasket in position on the 
flange manifold 

C

, ensuring screw holes are aligned 

correctly.

10. Position the flow disk assembly over the new flow disk 

lower gasket aligning the notched corner of the flow disk 
assembly 

J

 with the upper right-hand corner. Ensure 

screw holes are aligned correctly.

11. Apply a light coating of anti-seize compound (suitable for 

temperatures up to 200 °C [392 °F]) to the threads of 2 
(long) screws 

B

 and secure the flow disk assembly to 

the flange manifold. Tighten the screw to a torque of 
10 ±1 in.-lbs (1.1 Nm).

12. Apply 207 kPa (30 psig) air to air supply port 

K

 and check 

for leaks around the gasket and for no flow through 
aspirator 

L

. If leaks are detected, tighten the screws until 

all leaks are corrected. If flow is detected through the 
aspirator at room temperature, rebuild the flow disk 
assembly. 

13. Refer to Section 8.2.2, page 8 to refit the sensor cover.

14. Prepare the sensor for operation by reversing the 

shutdown / removal / isolation procedures – 
refer to Sections 5 and 4 (pages 4, 4).

WARNING – Bodily injury 

– Probe assemblies (standard and high temperature 

versions) contain fibrous material that can be a 
health hazard if airborne.

– Appropriate PPE, including mask, goggles and 

gloves must be worn when performing this 
procedure.

CAUTION – Damage to equipment 

Do not manually snap the flow disk. 
This will cause it to malfunction.

WARNING – Bodily injury 

– Do not use compressed air for any other 

purpose than that for which it is provided.

– Never direct a stream of compressed air 

towards your body or the body of any other 
person.

– When handling / connecting compressed air 

lines the correct safety procedures must be 
observed.

Summary of Contents for Endura AZ40

Page 1: ...nd 8 4 pages 4 4 6 and 12 Replacing sensor assembly flow disk Sections 4 5 6 and 8 5 pages 4 4 6 and 14 Kit contains Orifices gaskets seals This publication Tools required Transmitter door key supplie...

Page 2: ...contain fibrous material that can be a health hazard if airborne The material predominantly aluminosilicate refractory fibres CAS 142844 00 6 Refractory ceramic fibres RCF are classified as Category 1...

Page 3: ...sensor probe assembly must be mounted in accordance with the information supplied in Operating instruction OI AZ40 EN Suitable lifting equipment must be available when installing removing the sensor...

Page 4: ...the sensor air and test gas supplies at the process Referring to Fig 5 2 1 Disconnect air line B and test gas line C at sensor D DANGER Serious damage to health risk to life The transmitter must be i...

Page 5: ...embly B to mounting flange C Set items aside for re use 2 Carefully remove the sensor assembly and the attached probe including filter assembly from the process 3 Temporarily cover process opening unt...

Page 6: ...s kit component locations Fig 6 1 Orifice and seals kit component locations on sensor assembly Key A CO orifice see Section 7 page 7 B O2 orifice see Section 7 page 7 C CO heater block gaskets see Sec...

Page 7: ...a 1 4 in allen key wrench to secure CO orifice pipe plug A in place in the CO orifice chamber 9 Use a 5 32 in allen key wrench to secure O2 orifice D 1 notch in head in place in the O2 orifice chambe...

Page 8: ...flat bladed screwdriver to unscrew 4 captive cover retaining screws A and remove cover B Retain cover for re use 2 Use a medium flat bladed screwdriver to unscrew 2 captive cover retaining screws C an...

Page 9: ...crewdriver to depress the spring connector in each terminal and withdraw the wires from the terminal housing 3 Unscrew O2 sensor A from manifold block B using a 7 8 in spanner wrench and withdraw the...

Page 10: ...th attached washer K 14 Re fit COe block insulation A in the reverse order of removal 15 Apply a light coating of an anti seize compound suitable for temperatures up to 200 C 392 F to sensor clamp pla...

Page 11: ...Endura AZ40 Sensor assembly Orifice and seals kit INS ANAINST 007 EN Rev B 11 Fig 8 4 Replacing sensor assembly gaskets...

Page 12: ...ted correctly 7 Fit the nozzle B into the aspirator A ensuring it is seated correctly 8 Apply a light coating of an anti seize compound suitable for temperatures up to 200 C 392 F to the threaded port...

Page 13: ...Referring to Fig 8 6 If a recommended 2 stage filtration system A is fitted in the instrument air line replace the filters in accordance with manufacturers instructions Fig 8 5 Replacing Cleaning sen...

Page 14: ...rner of the flow disk assembly J with the upper right hand corner Ensure screw holes are aligned correctly 11 Apply a light coating of anti seize compound suitable for temperatures up to 200 C 392 F t...

Page 15: ...Endura AZ40 Sensor assembly Orifice and seals kit INS ANAINST 007 EN Rev B 15 Fig 8 7 Replacing sensor assembly flow disk...

Page 16: ...ts in this document and in the subject matter and illustrations contained therein Any reproduction disclosure to third parties or utilization of its contents in whole or in parts is forbidden without...

Reviews: