Endura AZ20 Oxygen Systems

Endura AZ20 – Pumped Reference Air Unit

3 Installation

4

INF09/027

Issue 3 

3.4 Accessing the Terminal Block

Referring to Fig. 3.3:

1. Loosen the 4 pump cover retaining screws 

A

.

2. Remove the pump cover 

B

.

3. Loosen cable gland 

C

 to prepare for the incoming

power supply cable. 

3.5 Electrical Connections

3.5.1 Preparation

Referring to Fig. 3.4:

1. Prepare the incoming power supply cable as follows:

a. cut the outer sheath back to leave 100 mm (4.0 in.)

sleeved wires 

A

b. cut the sleeved wires back to leave 5 mm (0.2 in.)

tails 

B

2. Pass the cable through cable gland 

C

.

3.5.2 Making Connections

Referring to Fig. 3.4, make mains connections as detailed in
steps 1 to 5. The mains power supply must be 115 or 230 V AC,
50 / 60 Hz depending on the pump model (refer to Fig. 3.3 and
Table 1.1, page 1).

1. Prepare the incoming power supply cable as follows:

a. cut the outer sheath back to leave 100 mm (4.0 in.)

sleeved wires 

A

b. cut the sleeved wires back to leave 5 mm (0.2 in.)

tails 

B

2. Pass the cable through cable gland 

C

.

3. Connect the neutral (blue wire 

D

) to the terminal block to

match the existing connection. 

4. Connect the earth wire 

E

 to the pump’s internal earth

connection.

5. Connect the live (brown) wire 

F

 to the terminal block to

match the existing live connection. 

6. Refit the pump cover by reversing the procedure in

Section 3.4.

Fig. 3.3 Accessing the Terminal Block

B

C

A

Pump Identification 

Label

Warning.

Ensure the live (brown) wire is protected 

with a 1.6 A Type F fuse.

Fig. 3.4 Electrical Connections

B

F

A

C

E

D