7 Operator Pages and Menus
56
OI/APA592–EN Rev. E
7.3.2 Calibrate
Used to calibrate the transmitter.
Calibrate Parameters
Parameter
Description
Tx
Type
PV Manual Cal
Reference procedure for calibration of the pH process value.
Options comprise:
1 Point Manual – see Table 7.2, page 58
2 Point Manual – see Table 7.3, page 59
Output Hold/Release – a quick access menu to hold or release the
current output
All
Auto Buffer Setup
pH
Buffer Type
Selects the type of buffer used:
User-defined
ABB
NIST
DIN 19266
Merck
US Tech
Buffer 1 Value
Enables selection of the
Buffer 1
value based on
Buffer Type
.
Enabled when
Buffer Type
is not
User Def
.
Buffer 2 Value
Enables selection of the
Buffer 2
value based on
Buffer Type
.
Enabled when
Buffer Type
is not
User Def
.
User def Soln.1
Enables entry of a user-defined Buffer Table 1 with up to 5 temperature and
pH values.
Enabled when
Buffer Type
=
User Def
.
User def Soln.2
Enables entry of a user-defined Buffer Table 2 with up to 5 temperature and
pH values.
Enabled when
Buffer Type
=
User Def
.
Auto Buffer Cal
Performs a two-point calibration based on
Auto Buffer Setup
and with
automatic sample stability checking – see Table 7.4, page 61.
If the sensor is removed from a hot or cold process allow it to normalize to
ambient temperature prior to calibration.
If the sensor cannot stabilize using automatic buffer calibration, perform a
manual calibration.
pH
2
Menu
Exit
Select
Calibrate