Endura APA592
pH / Redox (ORP) transmitter
4 Electrical Installation
OI/APA592–EN Rev. E
31
4.4.2 Sensors with Diagnostics Functions
Note.
Turn
On
diagnostics when connecting these sensors
Connection Information*
pH Sensor Cable Color
Transmitter
Terminal
Function
AP200
TBX5
Non ABB
4
Glass / Metal Sensing Electrode
Clear
Blue
Refer to sensor's
manual for wiring
colors and their
functions
5
Guard
Red
Yellow
6
Reference
Blue
Black
7
Solution Ground
Green/
Yellow
Green
8
Temperature Compensator
(If included with sensor)
Green
Red
9
Temperature Compensator
(If included with sensor)
White
White
10
Shield
–
Dark Green
11
Temperature Compensator
(If included with sensor)
Grey
–
Table 4.3 Sensors with Diagnostics Functions