Endura APA592
pH / Redox (ORP) transmitter
7 Operator Pages and Menus
OI/APA592–EN Rev. E
55
Temp. Comp. Type
(Temperature Compensation Type)
All
Manual
Manual
temperature compensation is enabled when the device does not
detect any temperature sensor connection either at power-up or upon
execution of
Run TC Recognition
. When
Temp Comp Type
is
Manual
, the
user configurable
Manual Temp
value is used for temperature compensation.
Auto
Enabled when a temperature sensor is provided with the probe (the
calculation is based on the Nernst equation – see Appendix A.3, page 98).
Enabled when a temperature sensor is detected by the device either at
power-up or when executing the
Run TC Recognition
routine is run (see
page 64).
Auto Solution
Auto Solution
temperature compensation can be selected when a
temperature sensor is connected and is recognized by the device. This
selection provides the ability to include a solution coefficient in the Nernst
equation – see Appendix A.3, page 98.
Manual Temp.
The value in this parameter is used for temperature compensation during
Manual
temperature compensation mode.
Adjustable range –20 to 150 °C (–4 to 302 °F). Default = 25 °C (77 °F).
Enabled when
Temp. Comp. Type
=
Manual.
All
pH Soln. Coeff.
Adjustable range –10 to 10. Default = 0 pH / 10 °C (50 °F).
Enabled when
Temp. Comp. Type
=
Auto.
pH
mV Soln. Coeff.
Adjustable range –20 to 20. Default = 0 mV / 10 °C (50 °F).
Enabled when
Temp. Comp. Type
=
Auto.
ORP /
pION,
Ion
Conc
Easy Setup Parameters
Parameter
Description
Tx
Type