110
EL3000
CONTINUOUS GAS ANALYZERS | CI/EL3000-EN REV. C
… 7 Operation
… LCD indicator
Status Icons
Icon
Description
An automatic calibration is in progress, see Operating
instruction.
The icon also appears in menu mode in the menu title line, see
A status message is active, see
"Diagnosis / error messages" in
The status signal ‘Maintenance required’ is active, see
"Diagnosis / error messages" in the operating instruction
The icon also appears in menu mode in the menu title line, see
page 110.
The status signal ‘Failure’ is active (see
messages" in the operating instruction
) or the maintenance
switch (see Operating instruction) is set to ‘On’. The icon is
flashing.
The icon also appears in menu mode in the menu title line, see
The configuration has been saved. The icon is flashing.
Do not turn off the power supply of the gas analyzer while the
icon is displayed!
Key functions in measurement mode
Button
Description
Toggle to display each individual measured value; in addition to
the number display, an analog bar with an indication of the
measuring range limits appears on this indicator.
Decrease or increase the contrast of the LCD display.
If a status message is active:
First, press the
button.
OK
Switch to menu mode (see 110).
If a status message is active
:
Press the button to display the message list (see "Diagnosis /
error messages" in the operating instruction).
Number of decimal places
When displaying the measured value in physical units (such as
ppm), the number of decimal places depends on how large the
span of the set measuring range is.
Measuring span
Decimal places
≤
0,05
5
≤
0,5
4
≤
5
3
≤
50
2
≤
500
1
> 500
0
The number of decimal places when setting the parameters is
the same as in the display in measuring mode.
LCD display in menu mode
Figure 51: Main menu
Structure of the menu
Each menu (see
menu contains a maximum of three menu items (‘3-tier menu’).
Each menu item is assigned to one of three buttons
,
and
,
therefore each menu item can be selected directly. The
button
is always used to return to the next higher menu.
The most commonly needed functions are arranged in the menu
in a way that they can be called up by pressing each of the keys
repeatedly:
‘
Operation /
Calibration /
Manual Calibration /
Zero point / Single point’
‘
Setup /
Calibration Data /
Test gas set points’
‘
Maintenance /
Diagnosis /
Device status /
Status
messages’