5
3.2.2
Panel Mounting – Fig. 3.5
3
MECHANICAL INSTALLATION
Locate instrument
in cut-out
Secure in panel using
four bolts, washers and nuts
Drill four suitable holes
Optional gasket
(see
Note 2
below)
4
5
3
Dimensions in inches (mm)
Cut hole in panel
(see
Note 1
below)
Mark four mounting holes
12.72 (323.08)
minimum
11.25
(285.8)
14.00 (355.6)
14.19
(360.4)
0.64 (16.25)
1.70
(43.2)
4 holes 0.281 dia.
or tap for
1
/
4
in.
thread
2
12.72
(323.1)
minimum
1
13.7 (348.0)
maximum
0.15 (3.8)
minimum
0.20
(5.0)
0.15 (3.8)
minimum
14.6
(371.0)
maximum
Ensure cut-out positioned centrally
between mounting holes
Minimum Cut-out Dimensions
Maximum Cut-out Dimensions
Notes.
1.
The instrument can be inserted into a panel cut-out
of any size between the minimum and maximum
dimensions illustrated, provided the cut-out is
positioned centrally relative to the fixing holes. If the
panel cut-out is larger than the maximum, a locally
manufactured adaptor plate will be required.
2.
If panel-mounting to NEMA 4X hosedown standard
is required, a continuous bead of suitable silicon
sealant
must
be applied beween the case flange
and the panel.
Do not
use the optional gasket.
Fig. 3.5 Panel Mounting
Summary of Contents for COMMANDER 1900 Series
Page 17: ...15 NOTES ...
Page 18: ...16 NOTES ...