4 Testing
4.1 Testing a Flow Metering System
The CalMaster2 can hold the test results of up to 100 flow meters.
The contrast of the LCD screen can be adjusted by pressing the
and
buttons.
To perform a test:
1.
Connect the CalMaster2 to the flow metering system – see Sections 3.2 to 3.4.
2.
Disconnect communications lead 1.
3.
Press the 'On' button. The initialization screen appears followed by a connection screen:
4.
The CalMaster2 attempts to connect to the flow measuring system. Select 'Ok'.
5.
Reconnect communications lead 1.
6.
Once the connection has been established a prompt appears to select whether the pipe is
full:
7.
Select
Yes
or
No
. A prompt appears to test the analog outputs:
Note.
If there is insufficient battery charge to completely a test, CalMaster2 does not start the
test. Recharge the CalMaster2 – see Section 6, page 40.
If the ambient temperature is greater than the specified operating temperature of 50°C
(122°F), the tests are failed.
Before using the CalMaster2 ensure that the flow metering system has been isolated
from any integrated system – for example, current loops and telemetry.
Note.
'Exp' is the number of days remaining before the licence expires.
Note.
When verifying an AquaMaster the pipe must always be full.
Connection