Field and Software Tools
CalMaster2 and CheckMaster
3 Installation
16
IM/VCM2
Issue 6
3.3.6 Fitting an Adaptor Box to the Sensor Cable
To fit an adaptor box to the sensor cable:
1.
Remove the cover of the adaptor box.
2.
Feed the sensor lead into the adaptor box
A
.
3.
Connect the sensor cable
B
.
4.
If required, provide environmental protection – see Section 3.3.7, page 16.
5.
Tighten the gland and replace the covers.
3.3.7 Adaptor Box Environmental Protection
Fig. 3.11 Fitting an Adaptor Box to the Sensor Cable
Fig. 3.12 Potting the Adaptor Box
Warning.
Potting materials are toxic – use suitable safety precautions.
Read the manufacturers’ instructions carefully before preparing the potting material.
Check all connections before potting – see Section 3.3.6, page 16.
Do not overfill or allow the potting material to come into contact with 'O' rings or grooves.
WABC2029
ABB
Belden 8777
Pin 1
Pin 2
Pin 3
Pin 4
Pin 5
Pin 6
Pin 7
Brown
Red
Orange
Yellow
Green
Blue
Violet
Black
Green
Black
Red
Sleeved Grounds
Black
White
B