15
Select Process Variable
Note.
No adjustment is available for the True Time Event Pen.
7
BASIC CONFIGURATION LEVEL
7.1.6
Set Up Relay Output
As detailed in
IM/C1900-PGC, Section 3.5,
with the following changes:
Information.
Relays can be energized by alarms, logic equation results, digital inputs, control and set point modes, totalizer
wrap signal (totalizer option on model 1953 only).
Select Relay Output
No setup is allowed for relay 1.1 which is assigned to the divert alarm, with polarity set as negative. All relays remain
de-energized for 10 seconds after instrument reset. This holds the instrument in failsafe divert mode.
Select Relay
Select required relay.
Relay Source
Select the source required to activate the selected relay – refer to
IM/C1900, Section 3.5,
Table 3.1
for source description.
rtd Or
(rtd overrange) is an additional source to those listed in Table 3.1.
Hot P (I/P1)
Cold P (I/P2)
Hot W (I/P3)
2
nd
RTD (Aux)
Model
Pen 1 (Red)
Pen 2 (Green)
No Pen
No Pen
1951
PV1
PV2
–
PV3
1952
PV3
PV2
PV1
PV4
1953
PV3
PV2
PV1
PV4
SELECt
----
PEN-AJ
ENbL-Y
PA PAS
-
N
C–PASS
-
Pen Adjust
Password frame to allow micrometer adjustment of the red pen.
N
If the password is not enabled (N selected), the micrometer adjustment value is not applied
to position of the red pen.
Y
If the password is enabled (Y selected), pen adjustment is included.
Select micrometer adjust code.
7.1.11
Scale Adjust
As detailed in
IM/C1900-PGC, Section 3.10
with the following changes:
7.1.7
Digital Inputs
As detailed in
IM/C1900-PGC, Section 3.8.
7.1.8
Set Up Digital Output
As detailed in
IM/C1900-PGC, Section 3.6
.
7.1.9
Set Up Analogue Output
As detailed in
IM/C1900-PGC, Section 3.7
.
7.1.10
Access Page
As detailed in
IM/C1900-PGC, Section 3.9
with the following additional frame which is displayed after the
Tune Password 2
frame:
SOUrCE
rtd Or
SELECt
rLY 2.1
3.1
4.1
Summary of Contents for C1900 Series
Page 19: ...Sales Software Service ...