48
…7
CALIBRATION
…7.1
Factory Settings
Calibrate Output 1
Note.
When adjusting the 4 and 20mA output values, the display reading is unimportant
and is used only to indicate that the output is changing when the
and
keys are
pressed.
See below.
Adjust 4mA
Set the milliammeter reading to 4mA.
Note.
The analog output range selected in
Configure Outputs
(see Section 5.5) does not
affect the reading.
Refer to Fig. 6.11 for connection details.
Adjust 20mA
Set the milliammeter reading to 20mA.
Note.
The analog output range selected in
Configure Outputs
(see Section 5.5) does not
affect the reading.
Refer to Fig. 6.11 for connection details.
Continued below.
Operating Page
– see Section 2.3.
Calibrate Output 2
Note.
Output 2 calibration is identical to Output 1 calibration.
Continued on next page.
Continued on next page.
Operating Page
– see Section 2.3.
Cal. Output 1
-----
O1: Adjust 4mA
16000
O1: Adjust 20mA
7200
Diss. Organics
Cal. Output 2
Cal. Output 1
Cal. Output 2
Cal. Output 2
-----
O1: Adjust 20mA
7200
Cal. Output 2
Option Board
Option Board
Diss. Organics