42/24-10 EN Rev. 9
Chapter 8: Gas Analyzer Calibration
8-C-9
Magnos206: Single-Point Calibration,
continued
Calibration data for manual calibration
Calibration method
Common single-point calibration
Calibration measurement range
0-25 Vol.-% O
2
Test gas concentration
20.96 Vol.-% O
2
Calibration data for automatic calibration
Calibration method
Common calibration (test gas)
Single zero calibration
always
Single span calibration
never
Common zero and span calibration
never
Test gas concentration zero gas
20.96 Vol.-% O
2
Calibration Data for an
Analyzer Module with
1 Sample Component
(Example:
Test Gas = Air)
Test gas concentration span gas
n.a.
Calibration data for manual calibration
Calibration method
Substitute gas calibration
Zero component
O
2
in N
2
Zero range
0-25 Vol.-% O
2
Span component
n.a.
Span range
n.a.
Test gas concentration zero gas
20.96 Vol.-% O
2
Test gas concentration span gas
n.a.
Calibration data for automatic calibration
Calibration method
Substitute gas calibration
1)
Single zero calibration
always
Single span calibration
never
Common zero and span calibration
never
Test gas concentration zero gas
20.96 Vol.-% O
2
Calibration Data for an
Analyzer Module with
> 1 Sample Component
(Example:
Test Gas = Air)
Test gas concentration span gas
n.a.
1) Settings of sample components and measurement ranges for zero and span
calibration are taken over from the calibration data for manual calibration.