42/24-10 EN Rev. 9
Chapter 8: Gas Analyzer Calibration
8-A-11
Calibration Methods,
continued
Overview
The following table summarizes the various calibration methods.
Quantity Calibration
SC MR Method
To configure ...
Calibrate ...
Calibration affects ...
1 1 Test
Gas /
Single
•
Zero
•
Span
individually for each sample
component and range
Measurement range
≥
1 > 1 Test Gas /
Common
Measurement ranges for
zero and span calibration
•
Zero in one measurement
range
•
Span in another range
for each sample component
All sample component
measurement ranges
> 1
≥
1 Substitute
Gas
Components and
measurement ranges for
zero and span calibration
•
Zero in one component
measurement range
•
Span in one range for
another component
for each detector
All detector components
and measurement ranges
SC = Sample and Substitute Gas Components
MR = Measurement Ranges per Component
Setting the
Calibration Method
(see Figure 8-A-7)
The calibration method can be set separately for manual, automatic and externally
controlled calibration.
For common and substitute gas calibration the sample ranges for start- and end-
point calibration of all three types of calibration control are adjusted jointly.
For substitute gas calibration the zero and span calibration components should
also be set.
Figure 8-A-7
Setting the
Calibration Method
Substitute
gas
Zero range
Span range
Common
calibration
Single
calibration
Zero range
Span range
Zero
component
Span
component
Test gas
Component
Calibration
method