Chapter 7: Gas Analyzer Configuration
7-D-11
Configuring Value Entries
Step Action
1
Select the menu item
Display
.
2
Start configuring the new display element with
NEW
.
3 Select
the
parameter
Page
.
The page overview is displayed.
Select the page on which the display element is to be displayed. Only
those pages can be selected with an assignment < 100 %, i.e., in
which there is at least one free position.
If the selected page is a:
system page, the position of the display element is set by the system;
it can only be changed using
Swap Display
(see
page 7-D-8).
4
user page,
the position must be configured.
5 Select
the
parameter
Position
.
The nine possible positions are graphically represented; free positions
are identified by the position number.
6
Select the position with the corresponding number key.
7 Select
the
parameter
Type
.
8 Select
display
type
Value entry
.
This creates a
constant
function block whose system-issued name,
‘Value page-position’, is displayed in the parameter
Source
. This
name cannot be changed here; it can only be changed by configuring
the function block (see Step 11).
9 Select
the
parameter
Config entry
and configure the other
parameters: entry range, decimal places, text and access level.
10
Enter the description of the display element in the parameter
Measuring point
.
11
Select the function block created in Step 8, enter the name and unit,
and link the function block to an application via its output 1 (for details,
see technical information “AO2000 Function Blocks – Descriptions and
Configuring” (document No. 30/24-200 DE)).
Configuring
Value Entries
12
Switch to measuring.
The newly configured display element is now shown in the display.
The description of the display element is shown above the display.
To the right of the display, the name and unit of the function block are
displayed that were entered in Step 11.