20
…10
MODBUS COILS AND REGISTERS
10.4
Multi-stream Phosphate Monitor, Model 8242/195
10.4.1 Coils
Coils used on Multi-stream monitors are identical to those used on single-stream versions – see Section 10.3.1.
10.4.2 Holding Registers
In addition to the Holding Registers listed in Section 10.3.2 above, Multi-stream monitors use the following Holding Registers:
Input Number
Read/Write
Description
Response/Entry
100
R
Phosphate Concentration,
Scaled 0 to 20.0 or 0 to 60.0 with decimal places
Stream 1
101
R
Phosphate Concentration
Scaled 0 to 20.0 or 0 to 60.0 with decimal places
Stream 2
102
R
Phosphate Concentration
Scaled 0 to 20.0 or 0 to 60.0 with decimal places
Stream 3
103
R
Phosphate Concentration,
Scaled 0 to 20.0 or 0 to 60.0 with decimal places
Stream 4
104
R
Phosphate Concentration
Scaled 0 to 20.0 or 0 to 60.0 with decimal places
Stream 5
105
R
Phosphate Concentration
Scaled 0 to 20.0 or 0 to 60.0 with decimal places
Stream 6
106
R/W
Stream 1 Decimal Point
0 = 1 d.p., 1 = 2 d.p.
107
R/W
Stream 2 Decimal Point
0 = 1 d.p., 1 = 2 d.p.
108
R/W
Stream 3 Decimal Point
0 = 1 d.p., 1 = 2 d.p.
109
R/W
Stream 4 Decimal Point
0 = 1 d.p., 1 = 2 d.p.
110
R/W
Stream 5 Decimal Point
0 = 1 d.p., 1 = 2 d.p.
111
R/W
Stream 6 Decimal Point
0 = 1 d.p., 1 = 2 d.p.
112
R/W
Alarm 1 Relay Setpoint
Scaled 0 to 20.0 or 0 to 60.0
113
R/W
Alarm 1 Relay On/Off Status
0=Off, 1=On
114
R/W
Alarm 1 Relay Action
0=Low, 1=High
115
R/W
Alarm 2 Relay Setpoint
Scaled 0 to 20.0 or 0 to 60.0
116
R/W
Alarm 2 Relay On/Off Status
0=Off, 1=On
117
R/W
Alarm 2 Relay Action
0=Low, 1=High
118
R/W
Alarm 3 Relay Setpoint
Scaled 0 to 20.0 or 0 to 60.0
119
R/W
Alarm 3 Relay On/Off Status
0=Off, 1=On
120
R/W
Alarm 3 Relay Action
0=Low, 1=High
121
R/W
Alarm 4 Relay Setpoint
Scaled 0 to 20.0 or 0 to 60.0
122
R/W
Alarm 4 Relay On/Off Status
0=Off, 1=On
123
R/W
Alarm 4 Relay Action
0=Low, 1=High
124
R/W
Alarm 5 Relay Setpoint
Scaled 0 to 20.0 or 0 to 60.0
125
R/W
Alarm 5 Relay On/Off Status
0=Off, 1=On
126
R/W
Alarm 5 Relay Action
0=Low, 1=High
127
R/W
Alarm 6 Relay Setpoint
Scaled 0 to 20.0 or 0 to 60.0
128
R/W
Alarm 6 Relay On/Off Status
0=Off, 1=On
129
R/W
Alarm 6 Relay Action
0=Low, 1=High
130
R/W
Relay Configuration
0=Concentration, 1=Out of Sample
131
R/W
Current Output 4 Range
Scaled 0 to 20.0 or 0 to 60.0
132
R/W
Current Output 5 Range
Scaled 0 to 20.0 or 0 to 60.0
133
R/W
Current Output 6 Range
Scaled 0 to 20.0 or 0 to 60.0
134
R/W
Current Output 1 Range
Scaled 0 to 20.0 or 0 to 60.0
135
R/W
Current Output 2 Range
Scaled 0 to 20.0 or 0 to 60.0
136
R/W
Current Output 3 Range
Scaled 0 to 20.0 or 0 to 60.0