17
10
MODBUS COILS AND REGISTERS…
10.3
Single-stream Phosphate Monitor, Model 8242/185
10.3.1
Coils
Input Number
Read/Write
Description
Response/Entry
001
R
Monitor Status
0 = Monitor In Service
1 = Monitor Not In Service
002
R
Monitor Calibration Status
0 = Monitor Not In Calibration
1 = Monitor In Calibration
003
R
Hold Mode
0 = Hold Mode Off
1 = Hold Mode On
004
R
Pump Status
0 = Pumps Off
1 = Pumps On
005
R
Upper Limit Control
0 = Control Tempe In Range
Temperature Status
1 = Control Tempe Out Of Range
006
R
Lower Limit Control
0 = Control Temperature – In Range
Temperature Status
1 = Control Temperature – Out Of Range
007
R
Reagent Status
0 = Monitor Not Out Of Reagent
1 = Monitor Out Of Reagent
008
R
Five-weekly Service Status
0 = Five-weekly Service Not Overdue
1 = Five-weekly Service Overdue
009
R
Yearly Service Status
0 = Yearly Service Not Overdue
1 = Yearly Service Overdue
010
R/W
Non-Vol Memory Mode
0 = Disable Write To Non-Volatile Memory
1 = Enable Write To Non-volatile Memory
011
R
Calibration Offset Alarm
0 = Calibration Offset Inside Of Limits
1 = Calibration Offset Outside Of Limits
012
R
Lower Calibration Slope Alarm
0 = Calibration Slope Inside Of Lower Limits
1 = Calibration Slope Outside Of Lower
013
R
Upper Calibration Slope Alarm
0 = Calibration Slope Inside Of Higher Limits
1 = Calibration Slope Outside Of Higher Limits
014
R
Sample Status
Monitor In Sample
Monitor Out Of Sample
Summary of Contents for 8241