Hydrazine Monitor
8 Troubleshooting
20
IM/7835–EN Rev. P
8 Troubleshooting
8.1 Diagnostics Classification Codes
Diagnostic messages and icons conforming to the NAMUR NE107 classification code are used to define information during operation
– a typical diagnostic message is shown in Fig. 8.1.
8.2 Diagnostics Messages
Fig. 8.1 Diagnostic Status Bar
CM30
80.0
20.0
Deg. C
Hydrazine µg/kg
In Recovery
In Recovery
Description of diagnostic
or alarm tag.
The highest priority diagnostic
or alarm is displayed.
Failure
Maintenance
Out of Specification
Check Function
NAMUR (NE107)
Status Icons
Operator
Page
High Process Alarm
Low Process Alarm
Icon
Message
Possible Cause
Suggested Action
Comms Failed
Power supply unit has failed or cable between
transmitter and sensor unit is damaged.
Check wiring between transmitter and sensor
unit.
Check voltage across red and black wires is
approximately 24 V. If problem not resolved
contact local service organization.
Pt1000 Failed
Pt1000 has failed or the cable from sensor
terminal box to temperature sensor is broken.
Check wiring inside sensor unit. If problem not
resolved contact local service organization.
ADC Failed
Temporary or permanent failure of analog to
digital converter on the sensor terminal board.
Cycle power to device. If problem persists
replace sensor terminal board/contact local
service organization.
Invalid Setting
Switch in sensor terminal box is set incorrectly.
Ensure switch adjacent to the terminals in the
sensor terminal box is in the up position.
In Recovery
After a calibration has been completed the
alarms and analog outputs are held for the
duration of the recovery period to allow the
sample to stabilize to the normal measurement.
Message disappears automatically once the
recovery period has expired. Alarms and
analog outputs become active.
Last Cal. Failed
Calibration has failed.
Sensor needs replacing.
Temperature Hot
The temperature of the sample water has
exceeded 55 °C (131 °F).
Hydrazine concentration is held automatically
at the last known value. Normal measurements
are resumed when the temperature falls below
55 °C (131 °F).
Temperature
Cold
The temperature of the sample water has fallen
below 5 °C (41 °F).
Hydrazine concentration is still displayed but a
fixed temperature compensation appropriate to
5 °C (41 °F) is applied.
Cal. Near Fail
Sensor approaching the end of its life and is likely
to fail soon.
Sensor (or system) requires attention soon.
Table 8.1 Diagnostic Messages