7835
Hydrazine Monitor
5 Commissioning
IM/7835–EN Rev. P
15
5.3.3 Sample Temperature
The temperature of the sample water is monitored continuously
using a Pt1000 resistance thermometer housed in the sensor
flowcell. Hydrazine measurement is compensated automatically
for variations in sample temperature and flow within the set
range – see Table 5.1, page 14.
If the temperature of the sample rises above 55 °C (131 °F), the
display reads 'hot' and the current outputs remains at the last
known values; returning to normal measurement when the
temperature falls below 55 °C (131 °F).
If the sample temperature falls below 5 °C (41 °F), hydrazine
concentration is still displayed but a fixed automatic temperature
compensation appropriate to 5 °C (41 °F) is applied.
The temperature is required to determine the necessary sample
flow from Table 5.1, page 14 when setting-up the monitor.
Fig. 5.1 Sample System During Normal Operation – Flow Diagram
Outlet Valve
Filler Cap
Outlet Valve
Filler Cap
Boyle-Mariotte
Tubes
Reagent
Dosing
Chamber
Microporous
Disc
Adjustable
Overflow
Tube
To Drain
Sensor
Solenoid Valve
Sample Inlet
Constant Head Unit
Standard Solution
Container
Reagent Container
Fig. 5.2 Sample System During Calibration Phase – Flow Diagram
Outlet Valve
Filler Cap
Outlet Valve
Filler Cap
Boyle-Mariotte
Tubes
Reagent
Dosing
Chamber
Microporous
Disc
Adjustable
Overflow
Tube
To Drain
Sensor
Solenoid Valve
Sample Inlet
Constant Head Unit
Standard Solution
Container
Reagent Container