7835
Hydrazine Monitor
6 Programming
18
IM/7835–EN Rev. P
6.2.3 Input/Output
6.2.4 Device Info
Configures analog outputs and relays.
Menu
Input/Output
Exit
Select
Analog Output 1(2)
The analog outputs can be set to retransmit an analog value with a configurable range from 0 to 20 mA.
Range
Sets the output range as one of: 0 to 10 mA, 0 to 20 mA or 4 to 20 mA.
Test
Performs a self-test against the selected mA range.
Alarm Relay 1(2)
Switches the alarm off.
Alarm activated on high process signal.
Alarm activated on low process signal.
Alarm Type
Off
High Process
Low Process
Alarm Trip
Sets a value at which the alarm is tripped.
Alarm Polarity
Sets the relay condition when an alarm is activated
Failsafe
The relay is de-energized (normally closed) when an alarm is activated.
Non-Failsafe
The relay is energized (normally open) when an alarm is activated.
Displays read-only factory-set parameters for the transmitter and sensor.
Trip
Point
Process
Variable
Alarm On
Alarm Off
Alarm On
Alarm
Off
High Process
Low Process
Menu
Device Info
Exit
Select
Transmitter
Serial Number
The transmitter’s serial number.
Date of Manufacture
The transmitter’s date of manufacturer.
Hardware Revision
The transmitter’s hardware revision.
Software Revision
The transmitter’s software revision.
Sensor
Sensor Type
The sensor type.
Serial Number
The sensor’s date of manufacturer.
Date of Manufacture
The sensor’s date of manufacturer.
Hardware Revision
The sensor’s hardware revision.
Software Revision
The sensor’s software revision.