background image

- 8 -

Fig. 7 - ConF METER menu

ConF

METER

ENTER

ConF

AUT O

ESC

ENTER

ESC

. . .   *

LOADING

NEXT

PREV

ENTER

NEXT

PREV

ConF

UPDA TE?

ConF

VIEW ?

ESC

ESC

ESC

0.000

ZERO

40.000

FULL   SC

KP A

ConF

MANUAL

OUTPUT

LINEAR

SQR

ESC

ENTER

NEXT

PREV

4000

ZERO

04000

04000

04000

00000

00000
ZERO

20000

FULL   SC

20000

FULL   SC

20000

20000

20000

20000

NEXT

PREV

NEXT

PREV

ENTER

ESC

ENTER

ESC

ENTER

ENTER

ESC

ENTER

UNITS

KP A

T ORR

ATM

MPA

IN H2O

KG / CM2

ENTER

NEXT  PREV

NEXT  PREV

ENTER

ENTER

NEXT

PREV

NEXT

PREV

NEXT

PREV

ENTER

ConF

CURRENT

ConF

UPDA TE?

ConF

PERCENT

ESC

ENTER

ENTER

ConF

PASSWD

NEXT

PREV

NEXT

PREV

ESC

ENTER

LINEAR

ConF

UPDA TE ?

SQR

ESC

ENTER

01234

PASSWD

00000

01000

01200

01230

UPDA TE?

ESC

ENTER

ESC

ENTER

ENTER

ENTER

NEXT

PREV

NEXT

PREV

ESC

ConF

OK

ENTER

ESC

NEXT

PREV

NEXT

PREV

The other options under ConF METER menu are:

ConF AUTO
By selecting this option, the CoMeter is automatically updated 
with the LRV, URV and Unit of the HART transmitter connected.
Before accepting the transmitter configuration by pressing 
ENTER at the request "ConF UPDATE?", it is possible to view 
the LRV (ZERO), the URV (FULL SC) and the UNIT.
If the output transfer function of the transmitter is not linear, 
ProMeter and CoMeter shows the message: ConF NO_LIN 
and the user cannot update the configuration.
It is necessary to change the output transfer function of the 
transmitter to linear.
See Fig. 7 - "ConF METER" menu, for ConF AUTO procedure.

ConF  MANUAL
The selection of MANUAL configuration allows the user 
to define manually the CoMeter configuration, i.e. define 
the LRV (ZERO), the URV (FULL SC), and the UNIT, as well 
as to decide for a LINEAR on SQR output function. LRV 
and URV can have a value between -99999 and +99999. 
Refer to Fig.7  - ConF METER menu for detail on the pro-
cedure.
For having the CoMeter to display the analog output current 
or the output percentage, select respectively:

ConF CURRENT   and   ConF PERCENT

CALIBRATION PROCEDURE  . . . 

UPDATE ZERO,

FULL-SCALE

AND UNIT

SET 4÷20 mA

INDICATION

SET 0÷100% INDICATION

(WITH OR WITHOUT

SQUARE ROOT)

UPDATE NEW

PASSWORD

COMETER only 

 

 

 

 

COMETER  and PROMETER

Summary of Contents for 600T Series

Page 1: ...1 Operating instruction IM 695FI_8 600T EN Series Model 695FI Field Indicator ASEA BROWN BOVERI...

Page 2: ...his manual is intended only to assist our customers in the efficient operation of our equipment Use of this manual for any other purpose is specifically prohibited and its contents are not to be repro...

Page 3: ...tes shown below The Nameplate provides information concerning the product code theinputsignal theproductserialnumber seeRef A Please refer to this number when making enquiries The Safety Marking plate...

Page 4: ...ter s terminal block fig 3 2 Junction box i e inserted in any point of the 4 to 20 mA line connecting the transmitter to its receiver power supply fig 4 In any cases the line resistance to the field i...

Page 5: ...EMOTE METER M Fig 3 Fig 4 To receiver power supply To receiver power supply 695FI as indicator only 695FI as junction box 2600T transmitter terminal block 2600T transmitter terminal block Fig 3a Fig 4...

Page 6: ...y in place screw on the cover and tighten WARNING ForHazardousAreasinstallations the connection of cables and conduits to the indicator shall be made in accordance with the requirements of the relevan...

Page 7: ...r itself and the transmitter can be configured ProMeter is only programmable In fact two are the main menu ConF METER and ConF XMTR ACCESS TO CONFIGURATION To enter these men for both indicators the k...

Page 8: ...itterconnected Before accepting the transmitter configuration by pressing ENTERattherequest ConFUPDATE itispossibletoview the LRV ZERO the URV FULL SC and the UNIT If the output transfer function of t...

Page 9: ...g the transmitter information See below a list of the available operation under the selected option CONF menu TRIM menu REVIEW menu PV menu Change LRV Reranging RERANG TAG 8 Primary variable PRIMARY C...

Page 10: ...UT 12000 OUT mA ENTER MANUAL ENTER SET 4mA ENTER 04 000 04 000 04 000 04 000 04 000 REF V AL 4 000 REF TX ENTER SET20 mA ENTER ENTER 20 000 20 000 20 000 20 000 20 000 REF V AL 20 000 REF TX ENTER ENT...

Page 11: ...ESC 0 000 LR V ESC 40 000 UR V ESC 40 000 LRL ESC 40 000 URL ESC 1 0000 DAMPING ESC OUTPUT LIN ESC ENTER ESC NEXT PREV NEXT PREV Fig 11 PV menu ENG UNITS PV ENTER PRIMAR Y ESC ENTER 10 sec 10 690 KP A...

Page 12: ...max span 16 mA EMI RFI Total effect 0 10 from 20 to 1000 MHz and for field strengths up to 10 V m when instrument is properly in stalled Input range 4 to 20 mA nominal Operating range 3 6 to 22 mA for...

Page 13: ...certificate number 01ATEX015 The meaning of ATEX code is as follows II Group for surface areas not mines 1 or 2 Category G Gas dangerous media D Dust dangerous media T50 C Maximum surface temperature...

Page 14: ...he number close to the CE marking of the indicator safety label identifies the Notified Body which carries out the Surveillance for the production of the indicator The other marking refers to the prot...

Page 15: ...surement Also liquid level and temperature I to P Converters and Field Indicators Flow Metering Magnetic Flowmeters Electromagnetic insertiontypeprobesandwatermeters Turbine Flowmeters Wedge Flow Elem...

Page 16: ...Ltd Howard Road St Neots Cambridgeshire PE19 3EU UK Tel 44 0 1480 475321 Fax 44 0 1480 217948 ABB Inc 125 E County Line Road Warminster PA 18974 USA Tel 1 215 674 6000 Fax 1 215 674 7183 ABB SpA Meas...

Reviews: