background image

2

4 69 0  S E R I E S

  |  4 6 70 TO 4 69 0 T U R B I D I T Y S Y S T EM U P G R A D E  |  I N F 1 1 /0 6 6 - EN R E V. A

3  EPROM replacement

Wall-mount transmitter

 WARNING

Remove all power from supply, relay and any powered 
control circuits and high common mode voltages before 
accessing the EPROM. 

 CAUTION

The transmitter is vulnerable to electrostatic damage. Wear 
an anti-static strap or dismantle the unit on an anti-static 
workbench. 

Isolate the transmitter from the power supply.

Referring to Figure 1:

Slide cover 

1

 down, pull out slightly and slide off.

Release 4 captive screws 

2

 and remove protection  

cover 

3

.

Remove 2 screw caps 

4

 (if fitted).

Remove 6 front panel screws 

5

 and remove front  

panel 

6

.

Figure 1  Removing the front cover – wall-mount transmitter

Referring to Figure 2:

Locate EPROM 

7

.

If required, remove personality module 

8

.

Remove EPROM using extraction tool.

Using EPROM insertion tool, position replacement EPROM 
carefully in socket ensuring notches in EPROM and socket 

9

 are aligned and that EPROM pins locate correctly in the 

holes in the socket.

Press EPROM fully home.

Refit personality module 

8

 (if removed at step 2).

8

7

9

Figure 2  Replacing the EPROM – wall-mount transmitter

Referring to Figure 1:

Refit front panel 

6

 and secure with 6 front panel  

screws 

5

.

Refit 2 screw caps 

4

 (if fitted).

Refit protection cover 

3

 and secure with 4 captive 

screws 

2

.

Refit cover 

1

.

Restore the power supply to the transmitter.

Configure the transmitter – see Section 4, page 3.

Reviews: